5.0 Operation
It is recommended that a pressure relief system should be used to avoid an over
pressurisation of the work tube.
Please note: A product should not be heated up if any valves that have been fitted are
closed to create a sealed volume. A sealed work tube should not be heated from cold
due to the pressure increase caused by the trapped air or gas expanding during the
heating process.
5.6
Gas Tightness
Work tubes of IAP material are impervious. Sillimanite may look similar but is porous.
Ensure that the correct tube material is in use before connecting and using gases other
than inert gases, such as nitrogen.
5.7 Running at Low Temperatures
The power limit may be adjusted to a low level in order to achieve better control when
operating the product at a low temperature. Before changing the power limit, record
the default settings for possible future use. Refer to the Power Settings section of this
manual for default power limits. If the product fails to reach the desired temperature,
refer to the Temperature Controller and Fault Analysis sections.