Fig 3 - Timer Modes
Key
T1
Temperature
T2
Time
SP
Setpoint
WSP
Working Setpoint (if setpoint ramp rate active)
AT
Actual Temperature
M245 Modes 2, 4 & 5
M13
Modes 1 & 3
TM
Timing
E
End
1
Mode 5
2
Reaching Temperature Modes 1 & 2
3
Modes 1 & 2
4
Modes 3 & 4 with setpoint ramp rate
5
Modes 3 & 4 with setpoint ramp rate off
6
Reaching temperature and continuing temperature Mode 5
7
Continuing at temperature modes 2 & 4, or cooling down modes 1 & 3
4.3.4
Stopping the Timer
To stop the timer at any time while it is operating, change the 'StAt User Calibration'
parameter to OFF. This is the same as reducing 'tmr' to zero. The controller then acts as
though it has reached the end of the time period.
45