2. Press the
Page
key until "Program Edit" is reached.
3. Press
Scroll
until the segment 6 dwell time is reached.
4. Adjust the dwell period to the calculated figure. The standard setting as supplied,
is 1 hour 40 minutes.
3.6
Power Adjustment
The product control system incorporates electronic power limiting. The power limit
parameter OP.Hi is accessible to the operator and can be used to adjust the product to
the actual supply voltage.
Some models covered by this manual are fitted with 208 V elements and are designed
for use over the range of voltages 200 V- 250 V; the power limit parameter is set
accordingly. These models may be relocated to a different voltage within the range: the
power limit should be reset to match the voltage.
To check whether the furnace is a 208 V model:
l
look at the OP.Hi parameter
l
make a note of the value and compare with the 9.0;
l
observe the rating label; if a range of voltages is stated, it is not a 208 V multi
voltage model;
l
if in doubt, check with Carbolite Gero – each product has its own record card.
Refer to the controller instructions for how to change the power limit.