Venus VIVA User Manual Effective Date: 28.07.13 P/N: DCVIROW3 Rev.01
10.
Energy Indicator Adjustment Button
: This button allows the operator to
determine the RF output of the Firm FX and Firm BX applicators once treatment
is initiated.
Energy is measured in percent of maximum (MP)
2
power output (120W), and
confined to 0-100%. Energy is adjusted by pressing the “+” and “-“ buttons
located adjacent to the Energy icon (marked 11A) and is made in one percent
increments.
Pressing on the "+" and "-" buttons continuously will adjust the
energy in ten percent increments.
11.
Treatment Time Adjustment Button
: This button allows the operator to
determine the treatment time, in minutes, by pressing the "+" and "-" buttons
located adjacent to the clock icon (marked 12B). Adjustment is made in one
minute increments and is confined to 0-30 minutes.
Treatment time can be adjusted during treatment.
Pressing the clock icon will set treatment time to zero.
12.
Interval Timer:
The Interval Timer allows the operator to divide the treatment
time into time segments. This feature can be used at operator's discretion, and
is convenient when a large treatment area is divided into smaller areas which
are then treated individually.
The
Interval Timer
begins countdown when treatment is initiated and beeps
when timer is zeroed. It then begins countdown again. This process repeats
itself as long as treatment is undergoing.
Interval Timer
is set between 0-30 minutes, and adjusted by pressing the “
+
” & “
-
“buttons located on the screen adjacent to the
Interval Timer
button (marked
13A). Adjustments are made in 15 seconds increments.
The
Countdown Timer
(marked 12B) begins countdown when treatment is
initiated and beeps when timer is zeroed.
Pressing the
Interval Timer
button will synchronize the Countdown Timer with
the Interval Timer.
13.
Start Button:
This button allows the operator to initiate treatment once
treatment parameters have been set.
Once the
Start
button has been pressed, it will illuminate in green, the system
will emit (MP)
2
energy through the selected applicator, and Treatment Time will
initiate countdown.
14.
Pause Button:
This button allows the operator to pause the treatment without
changing the selected treatment settings.
Once the
Pause
button has been pressed, it will illuminate in red, the system will
stop emitting (MP)
2
energy and treatment time countdown will temporarily stop.
Pressing the Start button will resume the treatment.
5-9