PTZ-BOX 5.0 User manual, document edition: 13
36
Functions
Entering data into the device
Configuration data can be entered into the PTZ-BOX 5.0 using device keyboard or by remote
transmission.
Remote configuration through digital communication channels (serial ports COM1, COM2, optical,
NFC and modem) can be performed with
PTZcom
software available from manufacturer’s website:
http://www.vemmtec.de/en
Configuration can be performed after login of user with adequate authorization level.
Passwords and privileges
Authorization system distinguishes 5 levels of privileges that may be used for logging into device and
2 (levels <=1) informational levels:
(9)
PRODUCER
(7)
METROLOGIST
(4)
ADMINISTRATOR
(3)
CUSTOMER
(2)
READER
(1)
BASIC
(0)
LOGOUT
Main features:
a)
Higher level has all privileges of lower level and additional privileges.
b)
Accounts of levels 2, 3, 4 are within protection range of hardware configuration lock
CFG
.
c)
Account of level 7 is within protection range of hardware metrological lock
MET
.
d)
Account of level 9 allows full, fixed access to device configuration (when
MET
is off).
Reserved only for manufacturer.
e)
Logging on levels 4 and 7 with hardware locks is possible only when switches
SecurLvlAdm
=1
or =3 and
SecurLvlMet
=3 (see below).
f)
Required privileges levels for each configurable parameter available in the device are listed in
document
“PTZ
-BOX 5.0 User d
ata structure”
.
In detail:
-
Levels 2-4 and 7 may use up to five fixed accounts (for each level, i.e. 201-205, 301-305, 401-
405, 701-705) built into device. By default, one account for each level is active: 201, 301, 401
and 701, with default password: 4096. Adding a user is performed by setting password
(different than 0) for that user. Users 401 and 701 cannot be removed.
-
Any modification of parameters requires login as selected user with correct password or
disabling one of hardware locks. After first login with use of keyboard - next modifications do
not require authorization. When entering sleep mode of the device, user is automatically
logged out. Next change of parameter requires to login again. Information about active user
logon is presented on status bar on main screen. Modification of parameters via transmission
requires to enter user identifier and correct password every time.