background image

 

1.  Introduction 

 

During  chemical-physical  treatments  of  waters  aimed  to  coagulate  and  precipitate  pollutants,  chemical  coagulants  are 
evaluated and their dosing optimized by jar tests. In relation to the variable composition of wastewaters the chemical dosing 
and  the  test  frequency  are  to  be  adapted  to  local  conditions.  JLT  Series  is  characterized  by  a  settable  speed  of  rotation, 
from  10  to  300  rpm,  with  the  real  speed  showed  on  the  relative  display.  The  steel  structure  epoxy  painted  is  studied  and 
tested in order to give the instrument an excellent resistance against the attack of chemical agents and corrosion in general. 
The stainless steel stirring rods are adjustable in height by a self-blocking chuck. The examined samples can be lighted by a 
back fluorescent lamp controlled by a switch located on front panel.  

2.  Assembly and installation 

 

Upon receipt and after having removed the packaging, please check the integrity of the instrument. The box includes: 

 

Flocculator 

 

Plug for power supply 

 

Switching VELP 100-240V/12V without plug 

 

Instruction manual 

2.1 Electrical connection 

Before connecting the instrument to the power supply, make sure that the values on the rating plate correspond to those of 
the power supply. Connect the unit to the power supply using the transformer supplied. 
Ensure that the socket and the relative cut-off device conform to current safety norms and easy to reach. 

2.2 Start-up 

Place the beakers containing the sample and then, set the speed and time 

using ↑ or ↓. 

3.  Operating controls 

 

The start-

up is empowered by the keys ↑ and ↓. t’s possible to select the speed from 10 to 300 rpm. The progression speed 

is increased by keeping one of the two keys pushed, in order to reach the desired value faster.  
It’s possible to select three different type of running by the Set key on the “Electronic Timer Window”: 
• Continuous 

 

 

• With timer (minutes base) 

 

 

• With timer (hours base) 

It’s possible to select the desired modality just by pressing the Set key and the display will show as follows: 
• Continuous way: “…..”  

• With timer (minutes): “000” 

 

 

• With timer (hours): “h00” 

In order to change the timer modality maintain  the Set key pushed for a few seconds. To set the desired time use the keys. 
If the time of running is not selected, the stirring is not active. 
The display shows the count down and simultaneously the third decimal point turns on to indicate that the count is operative. 
At the 

end, the instrument stops the stirring, the acoustic signal warns the end of the time and the display shows “End”. 

4.  Maintenance 

 

No routine or extraordinary maintenance is necessary apart from periodically cleaning the unit as described in this manual. 
In  compliance  with  the  product  guarantee  law,  repairs  to  our  units  must  be  carried  out  in  our  factory,  unless  previously 
agreed  otherwise  with  local  distributors.  The  instrument  must  be  transported  in  its  original  packaging  and  any  indications 
present  on  the  original  packaging  must  be  followed  (e.g.  palletized).  It  is  the  responsibility  of  the  user,  to  properly 
decontaminate the unit in case of hazardous substances remaining on the surface or interior of the device. If in doubt about 
the compatibility of a cleaning or decontamination product, contact the manufacturer or distributor. 

4.1 Cleaning 

Disconnect the unit from the power supply and use a cloth dampened with an non-inflammable non-aggressive detergent. 

5.  Technical data 

 

Power supply 

100-240 V / 50-60 Hz 

Power 

19 W 

Dimensions (WxHxD) 

655x404x296 mm (JLT4)         935x404x296 mm (JLT6) 

Weight (with plug) 

13.0 Kg (JLT4)                17.0 Kg (JLT6) 

Speed range 

10 ÷ 300 rpm, 1 rpm step 

Time range 

0 ÷ 999  minutes /  0 ÷ 99 hours or continuous  

Construction material 

Epoxy painted metal structure 

Environmental ambient temperature 

5 - 40 °C (41 - 104°F) 

Storage temperature range 

-

10… + 60 °C (14 - 140°F) 

Max humidity 

80% 

Pollution degree 

6.  Accessories / Spare parts 

 

A00001000 

Plastic Becher 1000 ml   

 

 

 

A00001001  

Glass Becher 1000 ml 

A00001002 

Transparent plastic Imhoff cone  

 

 

A00001003 

Glass graduated Imhoff cone 

A00001004 

Stand for 2 Imhoff cones 

 

 

 

10000235 

Foot  

EN 

Содержание JLT Series

Страница 1: ...areil comme d chet solide urbain conform ment la Directive 2002 96 CE No tirar el aparato en los desechos urbanos como exige la Directiva 2002 96 CE Dieses Ger t unterliegt der Richtlinie 2002 96 EG u...

Страница 2: ...pos el ctricos de medici n control y su uso en laboratorio Sicherheitsbestimmungen f r elektrische Mess Steuer Regel und Laborger te IEC EN 61010 1 Electrical equipment for laboratory use UL 61010 1 G...

Страница 3: ...to first appearing of flocs c evaluation of residual turbidity of supernatant after a determined sedimentation time by turbidimeter d measurement of electro kinetic potential of suspended particles o...

Страница 4: ...EN MARCHE 7 3 CONTR LES DES OP RATIONS 7 4 ENTRETIEN 7 4 1 NETTOYAGE 7 5 CARACT RISTIQUES TECHNIQUES 7 6 ACCESSOIRES PIECES DE RECHANGE 7 1 INTRODUCCI N 8 2 MONTAJE E INSTALACI N 8 2 1 CONEXI N A RED...

Страница 5: ...ssible to select the desired modality just by pressing the Set key and the display will show as follows Continuous way With timer minutes 000 With timer hours h00 In order to change the timer modality...

Страница 6: ...nte le frecce L agitazione non attiva se stato programmato un funzionamento temporizzato ed il tempo residuo 0 Il display visualizza costantemente il tempo residuo a scalare e contemporaneamente il te...

Страница 7: ...imal scintille en m me temps pour indiquer que le d compte est en cours A la fin du temps programm l agitation s arr te automatiquement et un signal sonore avertit l utilisateur que la proc dure est t...

Страница 8: ...de forma intermitente indicando que la cuenta del tiempo est activa A la fin del tiempo programado la agitaci n se detiene autom ticamente una se al ac stica se activa y la pantalla muestra end 4 Mant...

Страница 9: ...rt haben und die verbleibende Zeit 0 ist Das Display zeigt st ndig die verbleibende Zeit w hrend die dritte Dezimalstelle leuchtet in der intermittierende Form Am Ende der programmierten Zeit stoppt d...

Страница 10: ...araci n de conformidad Konformit tserkl rung We the manufacturer VELP Scientifica under our responsibility declare that the product is manufactured in conformity with the following standards Noi casa...

Страница 11: ...11...

Страница 12: ...yzers Digestion Units Distillation Units Solvent Extractors Fiber Analyzers Dietary Fiber Analyzers Oxidation Stability Reactor Consumables Analytical Instruments Analizzatori Elementari Digestori e M...

Отзывы: