5. External thermometers
VTF (only for
AREX 6 Connect
Pro)
➢
Move the main switch to off position
➢
Screw the threaded support rod into its seat on the back of the instrument
➢
Fasten the VTF thermoregulator onto the support rod
➢
Place the temperature probe in the flask
➢
Plug the VTF into the dedicated socket on the back of the instrument
➢
Switching on the instrument through the main switch, the probe led (A) turns on and display shows OFF
➢
Select operating temperature on VTF and after clicking temperature control knob, a dashed line (----)
appears on the display
NOTE:
while VTF is inserted, it is not possible to set any temperature turning the knob on the instrument.
The plate can heat until
the maximum temperature set in the full scale (see chapter 5, “Temp Limt”).
It is possible to set the timer only on the VTF.
External probe
➢
Move the main switch to off position
➢
Screw the threaded support rod into its seat on the back of the instrument
➢
Fasten the clamp onto the support rod
➢
Place the temperature probe in the flask
➢
Plug the probe into the dedicated socket on the back of the instrument
➢
Switching on the instrument through the main switch, the probe led (A) turns on and display shows OFF
➢
Select operating temperature using the temperature control knob
➢
Click the knob to start heating
➢
The display shows the real temperature obtained by the probe. Each 3 seconds the temperature set point
is shown on the display and the temperature set point led (C) turns on
6. Setting mode
Press both knobs for 3 seconds to enter into the setting mode when heating and stirring are off.
Clicking the Timer/autoreverse button (G) it is possible to pass from one of the following parameter to the next ones:
Parameter shown
Default
value
Range
Description
Display 1
Display
2
TENP
LINt
370
OFF - 50 -
370
Set full scale temperature:
➢
It limits the maximum value of temperature set point for the heating plate*
(from 50 to 370°C step 10°C)
➢
If “OFF” is set, it is not possible to perform any action with the
temperature control knob
SPEd
LINt
1700
100 - 1700
Set full scale speed:
➢
It limits the maximum value of speed set point for the motor (from 100 to
1700rpm, step 100rpm)
Strt
NOdE
StOP
StOP - run
Set starting mode:
➢
Stop: when the instrument is switched on through the main switch, OFF
is visualized on both displays
➢
Run: when the instrument is switched on, it restarts to work with the last
set points set
CtrL
tyPE
FInE
FInE
–
FASt
Set control temperature with external probe:
➢
Fine: slow temperature set point reaching. Overshoot optimized
➢
Fast: fast temperature set point reaching. High overshoot
PrOb
AlAr
On
On - OFF
Set external probe safety:
➢
If “on” it gives AL6 alarm when a slow temperature increase is detected
by the external probe. This function is active in case of:
-
Probe temperature < 50°C
-
T set point
– T probe > 5°C
➢
If “on” it gives AL7 alarm when a fast temperature decrease is detected
by the external probe.
PrOb
CAL
0.0
-10.0
÷
10.0
Alignment external probe:
➢
It allows external probe temperature alignment to a reference
thermometer
tINE
Strt
SnAP
SnAP - SEtP
Set start timer countdown:
➢
Snap: timer countdown starts as soon as one main function
(Temperature/Speed) is on
➢
SetP: timer countdown starts when temperature set point is reached
TINE
SPEd
On
On - OFF
Continue stirring after timer end:
➢
If
“on”, stirring continues after timer end