16

PersonalScope

TM

F  

Determining the frequency of a signal requires the
measurement of a period. The easiest way to do this
would be by placing the vertical markers either on
two consecutive peaks or two identical slopes of a
signal.

Setting up the  voltage per division

The user can enlarge or reduce the signal on the

screen vertically by adjusting the displayed voltage per
division. (V/div = voltage per division). The divisions can
be made visible through the 

Display

 key (see page 16).

Press the 

“t-V/div”

 key

Press the 

 or the 

 key to select the sensitivity : from

5mV to a maximum of 20V per division.

F  

50mV to 200V with X10 probe selection

Setting up the time base

Adjusting the time base will visualise more or

fewer periods of a signal (t/div = time per division). The
divisions can be visualised through the 

Display

 key (see

page 16).
Press the 

“t-V/div”

 key

Press the 

⇐ 

  the 

⇒  

key to set the time base between 20s

and 2µS per division.

ATTENTION :

 the positions 10µs, 4µs and 2µs only

operates correctly with periodic signals, for which the unit
will automatically select the ‘normal’ trigger mode. The

screen will be adjusted once the trigger level is reached.

NOTE: Use the minimum time base (2 microsec) as a
starting point when measuring a signal and select longer
time bases until the signal is displayed properly. Other-
wise the display may not correctly reflect the signal
under measurement due to aliasing.

Moving the signal in Y-position

Press the 

“Y-pos”

 key

Press the 

 key or the 

 key in order to move the signal

in the direction of the arrows. Prolonged pressing will make
the Y- position change faster. A small dot in the left-hand
corner of the screen will indicate the direction in which the
signal has moved. In this manner, the user will know in
which direction the signal was going when it went off-
screen.

t-V/div

t-V/div

Y-pos

OPERATION

Содержание Personal Scope HPS5

Страница 1: ......

Страница 2: ...2 PersonalScopeTM Velleman Legen Heirweg 33 9890 Gavere Belgium Internet Site http www velleman be E mail support velleman be HHPS5 1999 ED1...

Страница 3: ...Choice of input 11 Switching from one readout to another 12 Measuring probe adjustment 13 Auto setup function 13 Signal screen 14 Screen memory 15 Set up function keys 15 The markers 15 Setting up the...

Страница 4: ...na Increase the separation between the equipment and re ceiver Connect the equipment into an outlet on a circuit dif ferent from that to which the receiver is connected Consult the dealer or an experi...

Страница 5: ...It s ultra fast full auto setup function takes care of all the necessary settings and makes measuring waveforms very easy a feature that will be appreciated both by experts and novices Features High c...

Страница 6: ...o 400Vrms with X10 probe 2 5 accuracy Peak peak and DC range 0 1mV to 180V 1mV to 600V met x10 probe 2 accuracy Timebase range 20s 10s 5s 2s 1s 0 5s 0 2s 0 1s 50ms 20ms 10ms 5ms 2ms 1ms 0 5ms 0 2ms 0...

Страница 7: ...One should also refrain from measuring conductors or installations that use voltages that exceed 600Vrms above ground level CAT II indicates conformity for measurements of domestic installations The...

Страница 8: ...nal behind battery cover 5 Serial number 6 Display 7 Battery charge indication LED 8 Reset push button 9 Keyboard 9V DC I O 1 2 HPS5 mark 1 2 trig t V div Y pos probe x1 x10 Hold Display AC DC Auto Re...

Страница 9: ...prox 5 hours of autonomy while 800mA h types provide up to 8 hours of autonomy Attention Use a non regulated adapter of 9VDC that supplies a min of 300mA Observe the polarity and consult your retaile...

Страница 10: ...m right cor ner of your display when the batteries need replacement or charging Insufficient battery voltage may entail erro neous measuring results Survey of the indications on the screen Dot indicat...

Страница 11: ...by using the Reset function The unit will subsequently return to the manufacturer programmed setup Press the sunk Reset push button for at least 10 seconds F Do not use a sharp tool Adjusting the con...

Страница 12: ...ble periods one can also use the auto setup function dB measurement dB The measured signal is converted to dBm 0dB 0 775V Once more the user should make sure that at least one or two periods are displ...

Страница 13: ...10 measuring probes should be cali brated Read Setting up a X10 measuring probe Auto setup function The auto setup function is ideally suited for quick measurements as no manual setup has to be made a...

Страница 14: ...olt and time per division and Y position functions see further Signal screen The signal screen is an aid in displaying the divisions on your screen or in positioning movable markers on the screen Pres...

Страница 15: ...ments on a certain signal by using the four moveable markers This can be useful when measuring the interval between two points or the ampli tude of any given peak The following indications will appear...

Страница 16: ...ey see page 16 Press the t V div key Press the the key to set the time base between 20s and 2 S per division ATTENTION the positions 10 s 4 s and 2 s only operates correctly with periodic signals for...

Страница 17: ...el is reached Use this mode to detect e g a short once only voltage peak ATTENTION in the positions 10 s 4 s and 2 s the unit will automatically select the norm trigger mode The trigger level has to b...

Страница 18: ...he time per division to 0 2ms Select AC for the input Use the probe to perform measurements at the preselected point under the battery cover Adjust the trimmer of the measuring probe in order to obtai...

Страница 19: ...easure ments Take into account that the RMS readout only measures the AC voltage WARRANTY This product carries a one year warranty as far as the craftsmanship and possible flaws in the materials are c...

Страница 20: ...C5 100 GND R32 1K C25 100 L4 220 C26 100 AGND V2 V1 Personal Scope Power Supply Section C27 100 T1 BDP32 T2 BDP31 AGND GND D6 PRLL4001 D7 PRLL4001 T3 PMBT2222A T11 PMBT2907A T4 PMBT2222A T5 PMBT2222A...

Страница 21: ...AGND IN 3 IN 2 OUT 6 NC 8 OS NULL 1 OS NULL 5 VCC 7 VCC 4 IC4 LF357N 8 R7 47K RV1 22K C10 100n AGND V2 R36 4K7 1 C40 22p RV2 100 V2 C11 100n AGND R51 47 1 R53 180 1 R54 27 1 R45 2K4 1 R55 560 1 R52 4...

Страница 22: ...NC 17 NC 28 NC 40 VDD 12 VDD 35 IC7 PIC16C65A20 L 44 DGND X1 20Mc C41 22p C42 22p DGND DGND SW1 DGND AC DC ATT1 ATT2 ATT3 ATT4 ATT5 R8 10K C18 100n DGND C22 100 DGND DGND V1 V3 R24 10K DGND tVdiv Trig...

Страница 23: ...rweg 33 9890 Gavere Belgium Europe 32 0 9 3843611 http www velleman be HHPS5 1999 ED1 France VELLEMAN ELECTRONIQUE 33 0 320158615 Netherlands VELLEMAN COMPONENTS 31 0 765147563 USA VELLEMAN INC 1 817...

Отзывы: