Controls 

 

Oscilloscope mode 

  

  

VOLTS/DIV 

      

Selected value indicates the peak-to-peak voltage required to produce a peak-to-peak deflection of one 
major division on the screen. 
  

Coupling 

  

AC

: the input signal is capacitive coupled to the input amplifier/attenuator. Only the AC components are 

measured.   

DC

: the input signal is directly connected to the input amplifier/attenuator. Both AC and DC voltage are 

measured.   
  

TIME/DIV 

         

Selects the time setting for the beam to sweep one major division on the screen      
By selecting different TIME/DIV settings it is possible to 

zoom

 the frozen waveform on the screen.   

"Equivalent Time Sampling" 

ETS mode

 is in use on 50us/div, 20us/div and 10us/div ranges.   

Triggering is set ON to get stable waveform images.   
This mode is usable only for repetitive signals.   
There is no pre-trigger option in this mode.   
In this sampling mode the oscilloscope uses successive triggering occurrences to gather the data to 
construct a picture of a repetitive signal.   
  
  

TRIGGER On/Off

   

Selects free run mode or trigged mode.   
  

TRIGGER Level 

      

Selects the signal level at which the sweep is triggered.   
Triggering reference mark is displayed with the horizontal line on the left side of the screen.   
  

TRIGGER Edge

   

Selects the triggering slope:   
Up: Trigger occurs when triggering signal crosses trigger level in a positive going direction.   
Dn: Trigger occurs when triggering signal crosses trigger level in a negative going direction.   
  

TRIGGER Find

   

Resets the triggering x-position reference point. Triggering reference mark is displayed with the vertical line 
on the bottom of the screen.   
  

RUN

   

Selects recurrent display update mode (RUN). Pressing the button again freezes the display.   
  

SINGLE

   

When button is depressed and the trigger level is reached, refreshment of the display takes place only 
once.   

  
Persist 
  

When this button is down the scope captures many acquisitions of a signal to the screen. Record points 
accumulate until you release the button.   
  

Autoset

   

Automatic setup for the Volts/div, Time/div, and Trigger level to produce a stable waveform of usable size. 
The trigger will be set on if the wave aplitude on the screen is more than 0.5 divisions.   
The signal should be repetitive for proper autoset function: Amplitude 100mV to 30V; frequency above 50Hz; 
duty cycle greater than 10%.   
  

Содержание PCSU01

Страница 1: ...operating principle FFT Fast Fourier Transform o FFT resolution 512 lines transient recorder o timescale 20 ms div to 2000 s div o max recording time 9 4 h screen o automatic storage of data o record...

Страница 2: ...wer mains If necessary use a safety isolation transformer Warranty This product is guaranteed against defects in components and construction from the moment it is purchased and for a period of ONE YEA...

Страница 3: ...observe the max input rating of the scope Start measuring with trigger off 5 Press RUN 6 Choose the desired volts div setting or press Auto set 2 Choose the appropriate time div setting 7 To enable t...

Страница 4: ...the circuit under test observe the max input rating of the scope First observe the signal on the scope screen see previous page Check that the signal is not over the maximum of the screen Start the sp...

Страница 5: ...e circuit under test Select the appropriate volts div setting 1 Select the appropriate time division setting 3 Press RUN 2 to start recording Press RUN again to stop the measurement or use the Single...

Страница 6: ...vel Selects the signal level at which the sweep is triggered Triggering reference mark is displayed with the horizontal line on the left side of the screen TRIGGER Edge Selects the triggering slope Up...

Страница 7: ...the history menu The history menu options are Clear History Save History Open History Hide History Bar Show Hint Spectrum analyzer mode FREQ RANGE Sets the frequency range of the display Menu options...

Страница 8: ...is displayed in Run mode RMS Average Use this averaging mode to reduce signal fluctuations Vector Average Use this averaging mode to reduce random or uncorrelated noise in the synchronous signal you w...

Страница 9: ...rameters Software automatically calculates various voltage and time parameters of a signal such as DC mean amplitude rise time etc These parameters are displayed in a separate window Use the check box...

Страница 10: ...the text edit window is cleared and the selected waveform parameters data replaces the text Tip You can use Ctrl A Ctrl C and Ctrl V shortcuts in the text edit window To make the text transparent wit...