![Velleman DVM4x00 Series Скачать руководство пользователя страница 7](http://html1.mh-extra.com/html/velleman/dvm4x00-series/dvm4x00-series_user-manual_4076370007.webp)
CAT IV and up to 1000V in CAT III.
5.
Pollution degree
IEC 61010-1 specifies different types of pollution environments, for which different
protective measures are necessary to ensure safety. Harsher environments require
more protection, and the protection against the pollution which is to be found in a
certain environment depends mainly on the insulation and the enclosure properties.
The pollution degree rating of the DVM indicates in which environment the device
may be used.
Pollution
degree 1
No pollution or only dry, nonconductive pollution occurs. The pollution
has no influence. (only to be found in hermetically sealed enclosures)
Pollution
degree 2
Only nonconductive pollution occurs. Occasionally, temporary
conductivity caused by condensation is to be expected.(home and office
environments fall under this category)
Pollution
degree 3
Conductive pollution occurs, or dry nonconductive pollution occurs that
becomes conductive due to condensation that is to be expected.
(industrial environments and environments exposed to outside air - but
not in contact with precipitation)
Pollution
degree 4
The pollution generates persistent conductivity caused by conductive
dust or by rain or snow. (exposed outdoor environments and
environments where high humidity levels or high concentrations of fine
particles occur)
Warning:
This device was designed in accordance with EN 61010-1 pollution degree 2. This
implies that certain restrictions in use apply that are related to pollution which can
occur within the environment of use. Refer to the table above.
This device is only suitable for measurements in Pollution degree
class 2 environments.
6.
Description
Refer to the illustrations on pages 2 and 3 of this manual.
6.1
Multi-meter
1 LCD
7 COM terminal
2 function keys
8 PC link connector (DVM4100 only)
3 rotary switch
9 not used
4 VΩ
Hz terminal
10 foldable stand
5 µAmA terminal
11 battery compartment
6 10A terminal
6.2
Display
No. Symbol
Meaning
A
indicates negative readings
B
indicator for AC voltage or current
C
indicator for DC voltage or current
D
AUTO the meter is in the Auto range mode (automatically selects the
range with the best resolution)
E PC-LINK the meter is in the data transmission mode (DVM4100 only)
F
H
the meter is in Data Hold mode.
Содержание DVM4x00 Series
Страница 2: ...DVM4x00 V 0 2 13 02 2014 Velleman nv 2...