Press RIGHT ARROW KEY,
STEP 10, Input transient
Probe section
Connect the (
x10
) probe at
scope probe test output.
Adjust the probe calibration until
pass is indicated.
Press RIGHT ARROW KEY,
STEP 11, Input transient
Scope section
Connect the (
x10
) probe at
scope probe test output
Adjust CV3 until pass is indi-
cated
Press RIGHT ARROW KEY,
STEP 12, Calibration END
If all steps where correct “Pass”
should be indicated, but due to
possible manipulation Fail,
could be indicated even if all
tests where OK.
PRESS POWER
OFF TO END THE CALIBRATION