![background image](http://html2.mh-extra.com/html/vela/bariatric/bariatric_instructions-for-use-manual_4801814006.webp)
6
1.3. VELA TANGO WITH ELECTRIC HEIGHT ADJUSTMENT
1.4. LABELS USED ON THE CHAIR
Warning
Risk of shear and squeeze:
Keep hands and fingers away from moving parts
Icon
Read ”Instruction For Use” before use
Icon
Part used:
Type B cf. IEC60601-1
covering the following areas of the chair:
Labels
60 x 22,5 mm
The last six digits are the product ID number:
This label only applies to chairs
with electrical functions:
The chair is IP21 approved, which corresponds
to protection against water droplets, as well as
against touching dangerous parts with a finger.
Labels on the chair should be read at a distance of approx. 40 cm under normal
daylight conditions.
For labels related to the operation of the chair, please refer to section 2.0.
Follow instructions
for use
IP21
EN60601-1
Blade fuse: 10A/32V,
0.5A
Input:
24V
Duty cycle: 10%,
max. 2min/18min Storage limits
-20°C
60°C
1.3.1. POWER INPUT
Voltage: 24Vdc
Current: 0.5A
Power: 12VA
1.3.2. EMISSION
A VELA Medical chair with electrical functions has been tested and complies
with current requirements regarding electromagnetic radiation (EMC require-
ments). The chair is tested to the following specifications:
::
Emission class CISPR 11, Group 1, Class B
See more details at www.vela-medical.com
UDI:
(XX) XXXXXXXXXXXXXX (XX) XXXXXX
<Model name>
<Model ref. no.>
<Manufacturer>
<Adress>
<Webpage>
<Safe working load>
<Production date/order no.>
MEDICO