25
6 Set up with VEGADIS 62
VEGAPULS WL 61 • 4 … 20 mA/HART - two-wire
38061-EN-121011
6 Set up with VEGADIS 62
6.1 Connection
The VEGADIS 62 is an indicating and adjustment unit without external
energy for looping into 4 … 20 mA/HART circuits.
The parameter adjustment of the sensor is carried out via HART com-
munication. During the parameter adjustment, the VEGADIS 62 acts
as a Secondary Master with respect to the sensor.
~
=
1
2
4
3
Fig. 24: Connection of VEGADIS 62 to the sensor
1 Sensor
2 VEGADIS 62
3 HART resistance > 150 Ω (necessary with low impedance power supply)
4 Voltage supply/Processing
The following adjustment volume of the connected HART sensor is
available:
•
Min./Max. adjustment
•
zero/span adjustment (live adjustment)
•
Damping
6.2 Adjust the sensor
Proceed as follows for the min./max. adjustment of the sensor:
1. Press "
OK
" to reach the adjustment menu.
2. Select the submenu "
Measurement
" and confirm with "
OK
".
3. Move to the menu item "
Unit
". There the instrument unit of the
sensor is displayed, for example "
m
".
4. Move to the menu item "
MB begin
", there the max. measuring
distance is displayed, for example the default setting 15 m.