5 Set up with the display and adjustment module PLICSCOM

VEGAPULS 68 • 4 … 20 mA/HART - four-wire

51905-EN-180222

5.2  Setup steps

1.  Go to the menu "

Basic adjustment

" via the display and adjust-

ment module.

Basic adjustment
Display
Diagnostics
Service
Info

2.  Carry out the adjustment in the menu items "

Min. adjustment

" and 

"

Max. adjustment

".

Max. adjustment

100.00 %

=

1.000 m(d)

2.000 m(d)

Min. adjustment

0.00 %

=

5.000 m(d)

4.000 m(d)

3.  Select in the menu item "

Medium

" the medium of your applica-

tion, for example "

Powder/Dust

".

Medium

Bulk solid

4.  Select in the menu item "

Application

" the vessel and the vessel 

form, for example, "

Silo

".

Vessel form

Silo

Since the radar sensor is a distance measuring instrument, it is the 

distance from the sensor to the product surface that is measured. To 

indicate the actual level, the measured distance must be assigned to 

a certain height percentage.
To perform the adjustment, enter the distance with full and empty ves-

sel, see the following example:

Set parameters

Setup - Adjustment

Содержание VEGAPULS 68

Страница 1: ...Quick setup guide Radar sensor for continuous level measurement of bulk solids VEGAPULS 68 4 20 mA HART four wire Document ID 51905...

Страница 2: ...ting instructions 8 4 Connecting to power supply 10 4 1 Connection procedure 10 4 2 Wiring plan double chamber housing 11 5 Set up with the display and adjustment module PLICSCOM 12 5 1 Insert display...

Страница 3: ...em components through incorrect mounting or adjustment Thus dam age to property to persons or environmental contamination can be caused Also the protective characteristics of the instrument can be inf...

Страница 4: ...netic interference from electrical conductors and radiated emissions must be taken into account as is usual with a class A instrument according to EN 61326 1 If the instrument is used in a different e...

Страница 5: ...Take note of the respective operating regulations This device may not cause interference and The device must be resistant to interference signals including such that may cause undesired operating stat...

Страница 6: ...orks With the serial number you can access the delivery data of the instru ment via www vega com VEGA Tools and Instrument search You can find the serial number on the inside of the instrument as well...

Страница 7: ...it against falling off 4 Retighten the antenna with hexagon screws to the antenna socket torque max 10 Nm 7 5 lbf ft Note The VEGAPULS 68 with rinsing air connection or antenna exten sion is provided...

Страница 8: ...maged Note A damaged O ring seal must be replaced FKM article no 2 28248 FFKM Kalrez 6375 article no 2 27351 8 Remount the parabolic antenna 4 9 Tighten compression nut 1 with SW 41 torque max 50 Nm 1...

Страница 9: ...ire 51905 EN 180222 ca 10 mm 200 mm 7 87 Fig 3 Distance of the antenna to the vessel wall vessel ceiling 2 Note min socket diameter depending on the socket length 3 Note the instructions for sealing F...

Страница 10: ...sensor through the cable entry 6 Lift the opening levers of the terminals with a screwdriver see following illustration 7 Insert the wire ends into the open terminals according to the wir ing plan Fi...

Страница 11: ...51905 EN 180222 4 2 Wiring plan double chamber housing The following illustrations apply to the non Ex as well as to the Ex ia version 4 20 mA PE L N L1 N GND 1 2 1 2 3 4 4 20mA IS Fig 5 Wiring plan d...

Страница 12: ...on the electronics you can choose any one of four different posi tions each displaced by 90 3 Press the display and adjustment module onto the electronics and turn it to the right until it snaps in 4...

Страница 13: ...000 m d 4 000 m d 3 Select in the menu item Medium the medium of your applica tion for example Powder Dust Medium Bulk solid 4 Select in the menu item Application the vessel and the vessel form for e...

Страница 14: ...ng the product level These settings can be made ahead of time without the instrument having to be installed High sockets or vessel installations such as e g struts or agitators as well as buildup and...

Страница 15: ...in this area 5 3 Menu schematic Information Depending on the version and application the light coloured menu windows are not always available or offer nor selection possibility Basic adjustment Basic...

Страница 16: ...play Diagnostics Service Info False signal suppression 4 1 Change now Extended setting 4 2 None Quick level change Current output 4 3 Characteristic 4 20 mA Fail mode 3 6 mA min current 4 mA max curre...

Страница 17: ...supply Electromechanical data version IP 66 IP 67 Cable entry Double chamber housing 1 x cable gland M20 x 1 5 cable 5 9 mm 1 x blind plug M20 x 1 5 plug M12 x 1 for VEGADIS 61 optional or 1 x closin...

Страница 18: ...18 Notes VEGAPULS 68 4 20 mA HART four wire 51905 EN 180222...

Страница 19: ...19 Notes VEGAPULS 68 4 20 mA HART four wire 51905 EN 180222...

Страница 20: ...of delivery application practical use and operat ing conditions of the sensors and processing systems correspond to the information available at the time of printing Subject to change without prior n...

Отзывы: