Vega VEGADIF 85 Скачать руководство пользователя страница 6

6

3 Mounting

VEGADIF 85 • Profibus PA

53575-EN-230814

3 Mounting

3.1  General instructions for use of the instrument

Note:

For safety reasons, the instrument must only be operated within the 

permissible process conditions. You can find detailed information on 

the process conditions in chapter " 

Technical data

" of the operating 

instructions or on the type label. 
Hence make sure before mounting that all parts of the instrument ex-

posed to the process are suitable for the existing process conditions.
These are mainly:

• 

Active measuring component

• 

Process fitting

• 

Process seal

Process conditions in particular are:

• 

Process pressure

• 

Process temperature

• 

Chemical properties of the medium

• 

Abrasion and mechanical influences

Protect your instrument against moisture ingress through the following 

measures:

• 

Use a suitable connection cable (see chapter " 

Connecting to 

power supply

") 

• 

Tighten the cable gland or plug connector

• 

Lead the connection cable downward in front of the cable entry or 

plug connector

This applies mainly to outdoor installations, in areas where high 

humidity is expected (e.g. through cleaning processes) and on cooled 

or heated vessels.

Note:

Make sure that during installation or maintenance no moisture or dirt 

can get inside the instrument.
To maintain the housing protection, make sure that the housing lid is 

closed during operation and locked, if necessary.

The ventilation for the electronics housing is realised via a filter ele-

ment in the vicinity of the cable glands.

Process conditions

Protection against mois-

ture

Ventilation

Содержание VEGADIF 85

Страница 1: ...Quick setup guide Differential pressure transmitter with metallic measuring cell VEGADIF 85 Profibus PA Document ID 53575...

Страница 2: ...t display and adjustment module 10 5 2 Parameter adjustment Quick setup 11 5 3 Parameter adjustment Extended adjustment 15 5 4 Menu overview 15 6 Set up with smartphone tablet PC notebook via Bluetoot...

Страница 3: ...ns This is a state of the art instrument complying with all prevailing regulations and directives The instrument must only be operated in a technically flawless and reliable condition The operating co...

Страница 4: ...r fault information from measuring transduc ers NE 53 Compatibility of field devices and display adjustment components NE 107 Self monitoring and diagnosis of field devices For further information see...

Страница 5: ...nical data Serial number of the instrument QR code for device identification Numerical code for Bluetooth access optional Manufacturer information To find order data documents or software related to y...

Страница 6: ...he medium Abrasion and mechanical influences Protect your instrument against moisture ingress through the following measures Use a suitable connection cable see chapter Connecting to power supply Tigh...

Страница 7: ...el single chamber precision casting 2 Aluminium single chamber 3 Stainless steel single chamber electropolished 4 Plastic double chamber 5 Aluminium stainless steel double chamber housing precision ca...

Страница 8: ...of the cable gland and remove blind plug 4 Remove approx 10 cm 4 in of the cable mantle strip approx 1 cm 0 4 in of insulation from the ends of the individual wires 5 Insert the cable into the sensor...

Страница 9: ...connection compartment single chamber housing 1 Voltage supply signal output 2 For display and adjustment module or interface adapter 3 Selection switch for instrument address 4 For external display a...

Страница 10: ...90 It is not necessary to interrupt the power supply Proceed as follows 1 Unscrew the housing lid 2 Place the display and adjustment module on the electronics in the desired position and turn it to t...

Страница 11: ...y adapt the sensor to the application select the menu item Quick setup in the start graphic on the display and adjustment module Carry out the following steps in the below sequence You can find Extend...

Страница 12: ...set on the measured value Min adjustment In this menu item you carry out the min adjustment for level Enter the percentage value and the corresponding pressure value for the min level Max adjustment I...

Страница 13: ...value and the corresponding density value for the min density Max adjustment In this menu item you carry out the max adjustment for density Enter the percentage value and the corresponding density va...

Страница 14: ...variable correctly an allocation of the output signal to 0 and 100 must be carried out adjustment When using the application Level the hydrostatic pressure e g with full and empty vessel is entered as...

Страница 15: ...AI FB 1 Channel scaling damping Display Settings e g for language measured value display lighting Diagnosis Information e g on instrument status peak indicator measurement reliability AI FB 1 simulati...

Страница 16: ...tment Released Menu item Default value Menu language Order specific Displayed value 1 Signal output in Displayed value 2 Measuring cell temperature in C Backlight Switched on Menu item Parameter Defau...

Страница 17: ...t Unit kg s Adjustment 0 corresponds to 0 kg s 100 corresponds to 1 kg s Menu item Parameter Device name Device name Instrument version Hardware and software version Factory calibration date Date Prof...

Страница 18: ...luetooth 1 Switch On Bluetooth active Off Bluetooth not active The security concept of Bluetooth operation absolutely requires that the default setting of the sensor PIN be changed This prevents unau...

Страница 19: ...art PACTware and the VEGA project assistant Select the device search via Bluetooth and start the search function The device auto matically searches for Bluetooth capable devices in the vicinity The me...

Страница 20: ...20 6 Set up with smartphone tablet PC notebook via Bluetooth VEGADIF 85 Profibus PA 53575 EN 230814 Fig 9 Example of an app view Setup sensor adjustment App view...

Страница 21: ...d Ex d instruments 12 mA 0 5 mA Electromechanical data version IP66 IP67 and IP66 IP68 0 2 bar 1 Options of the cable entry Cable entry M20 x 1 5 NPT Cable gland M20 x 1 5 NPT cable see below table Bl...

Страница 22: ...22 Notes VEGADIF 85 Profibus PA 53575 EN 230814...

Страница 23: ...23 Notes VEGADIF 85 Profibus PA 53575 EN 230814...

Страница 24: ...ing scope of delivery application practical use and operat ing conditions of the sensors and processing systems correspond to the information available at the time of printing Subject to change withou...

Отзывы: