5 Set up with the display and adjustment module
VEGADIF 85 • Profibus PA
53575-EN-230814
Zero adjustment
In this menu item you carry out the Zero adjustment for the differential
pressure.
Enter the corresponding pressure value for 0 %.
Span adjustment
In this menu item you carry out the Span adjustment for the differen-
tial pressure
Enter the corresponding pressure value for 100 %.
Distance
In this menu item, you enter the installation distance between Primary
and Secondary sensor.
Min. adjustment
In this menu item you carry out the min. adjustment for density.
Enter the percentage value and the corresponding density value for
the min. density.
Max. adjustment
In this menu item you carry out the max. adjustment for density.
Enter the percentage value and the corresponding density value for
the max. density.
Distance
In this menu item, you enter the installation distance between Primary
and Secondary sensor.
Min. adjustment
In this menu item, you carry out the adjustment for the min. height of
the interface.
Enter the percentage value and the corresponding height of the
interface.
Max. adjustment
In this menu item, you carry out the adjustment for the max. height of
the interface.
Enter the percentage value and the corresponding height of the
interface.
Quick setup - Differential
pressure measurement
Quick setup - Density
measurement
Quick setup - Interface
measurement