5 Connecting to power supply
VEGACAL 67 • For connection to a controller
31761-EN-200317
5 Connecting to power supply
5.1 Preparing the connection
Always keep in mind the following safety instructions:
•
Connect only in the complete absence of line voltage
•
If overvoltage surges are expected, overvoltage arresters should
be installed
Tip:
We recommend using VEGA overvoltage arresters B63-48 and
ÜSB 62-36G.X.
Power supply and current signal are carried on the same two-wire
cable. The voltage supply range can differ depending on the instru
-
ment version.
The data for power supply are specified in chapter "
Technical data
".
Provide a reliable separation between the supply circuit and the
mains circuits according to DIN VDE 0106 part 101. The VEGA power
supply units VEGATRENN 149AEx, VEGASTAB 690, VEGADIS 371
as well as all VEGAMETs meet this requirement.
Keep in mind the following additional factors that influence the operat
-
ing voltage:
•
Output voltage of the power supply unit can be lower under nomi-
nal load (with a sensor current of 20.5 mA resp. 22 mA in case of
fault message)
•
Influence of additional instruments in the circuit (see load values in
chapter "
Technical data
")
The instrument is connected with standard two-wire cable without
shielding. If electromagnetic interference is expected which is above
the test values of EN 61326 for industrial areas, shielded cable should
be used.
Make sure that the cable used has the required temperature resist-
ance and fire safety for max. occurring ambient temperature
Use cable with round cross-section. A cable outer diameter of
5 … 9 mm (0.2 … 0.35 in) ensures the seal effect of the cable gland.
If you are using cable with a different diameter or cross-section,
exchange the seal or use a suitable cable gland.
If shielded cable is required, connect the cable screen on both ends
to ground potential. In the sensor, the screen must be connected
directly to the internal ground terminal. The ground terminal on the
outside of the housing must be connected to the potential equalisa-
tion (low impedance).
If potential equalisation currents are expected, the connection on the
processing side must be made via a ceramic capacitor (e. g. 1 nF,
1500 V). The low-frequency potential equalisation currents are thus
suppressed, but the protective effect against high frequency interfer
-
ence signals remains.
Note safety instructions
Voltage supply
Connection cable
Cable screening and
grounding
Содержание VEGACAL 67
Страница 3: ...3 Contents VEGACAL 67 For connection to a controller 31761 EN 200317 Editing status 2020 03 03...
Страница 33: ...33 Notes VEGACAL 67 For connection to a controller 31761 EN 200317...
Страница 34: ...34 Notes VEGACAL 67 For connection to a controller 31761 EN 200317...
Страница 35: ...35 Notes VEGACAL 67 For connection to a controller 31761 EN 200317...