background image

5

2 Mounting

Screening tube adapter • for VEGACAL 63, 64/VEGACAP 63

34849-EN-130916

Caution:

Check the insulation of the probe for deep scratches.

To ensure the seal effect of the screening tube adapter, the isolation 

of the probe must be undamaged in the area of the lower screening 

tube adapter end.
1.

 

Loosen and remove (SW

 

24) pressure screw (5)

2.

 

Remove the two O-rings (4) from the screening tube

3.

 

Place the process seal (2) on the thread of the probe

4.

 

Push the rod of the probe (1) through the screening tube adapter 

(3)

5.

 

Screw the screening tube adapter (3) onto the thread of the probe 

and tighten with 100

 

Nm (74

 

lbf

 

ft) (SW

 

46)

6.

 

If it is harmless to the process, grease the two O-rings with an 

acid-free grease so that the rod of the probe slides more easily

7.

 

Slide the two O-rings (4) over the rod of the probe

8.

 

Slide the pressure screw (5) of the screening tube adapter onto 

the probe

9.

 

Screw pressure screw (5) with the two O-rings into the screening 

tube adapter and tighten it to the stop (SW 

24)

The mounting of the screening tube adapter is now finished. The 

probe can now be installed.

You can find the mounting procedure in the operating instructions of 

the probe.

Note:

The VEGACAP probe must be re-compensated after mounting the 

screening tube adapter. Use the compensation key on the electronics 

module. You can find the instructions for this in the operating instruc

-

tions of the probe.

Содержание VEGACAL 63

Страница 1: ...Supplementary instructions Screening tube adapter for VEGACAL 63 64 VEGACAP 63 Rod 16 mm Document ID 34849...

Страница 2: ...ning vacuum tight Screening against high capacitances caused by installation condi tions This can occur e g in long sockets In addition the screen ing tube is vacuum tight In case of strong condensati...

Страница 3: ...g cone PTFE Caution Check the insulation of the probe for deep scratches To ensure the seal effect of the screening tube adapter the isolation of the probe must be undamaged in the area of the lower s...

Страница 4: ...nting procedure in the operating instructions of the probe Note The VEGACAP probe must be re compensated after mounting the screening tube adapter Use the compensation key on the electronics module Yo...

Страница 5: ...with 100 Nm 74 lbf ft SW 46 6 If it is harmless to the process grease the two O rings with an acid free grease so that the rod of the probe slides more easily 7 Slide the two O rings 4 over the rod o...

Страница 6: ...L 63 VEGACAL 64 require ment 16 mm 0 63 in Process fittings Inner thread G1 A pipe thread cylindrical ISO 228 T1 Outer thread G1 A pipe thread cylindrical ISO 228 T1 Weight Basic weight approx 500 g 1...

Страница 7: ...nsions Screening tube adapter against condensation for VEGACAL VEGACAP rod 16 mm 0 63 in 27 mm 1 1 16 G1 1 2 G1 SW 64 L1 16 mm 5 8 24 mm 15 16 Fig 3 Screening tube adapter against condensation for VEG...

Страница 8: ...130916 Screening tube adapter vacuum tight for VEGACAL VEGACAP rod 16 mm 0 63 in 27 mm 1 1 16 L1 G1 1 2 G1 SW 64 16 mm 5 8 24 mm 15 16 Fig 4 Screening tube adapter vacuum tight for VEGACAL VEGACAP ro...

Страница 9: ...9 Notes Screening tube adapter for VEGACAL 63 64 VEGACAP 63 34849 EN 130916...

Страница 10: ...10 Notes Screening tube adapter for VEGACAL 63 64 VEGACAP 63 34849 EN 130916...

Страница 11: ...11 Notes Screening tube adapter for VEGACAL 63 64 VEGACAP 63 34849 EN 130916...

Страница 12: ...of delivery application practical use and operat ing conditions of the sensors and processing systems correspond to the information available at the time of printing Subject to change without prior n...

Отзывы: