background image

6

4 Mounting

Electronics module • VEGACAL series 60

30531-EN-170531

4 Mounting

4.1  General instructions

If the electronics module is defective, it can be replaced by the user.
In Ex applications, only instruments and electronics modules with ap-

propriate Ex approval may be used.
If there is no electronics module available on site, one can be ordered 

from the VEGA agency serving you.

The new electronics module must be loaded with the order and fac-

tory data of the sensor. These are the options:

• 

At the factory by VEGA

• 

Or on site by the user

Information:

When loading on site, the order data must first be downloaded from 

the Internet (see "

Setup

").

In both cases, the sensor serial number is necessary. The serial num-

bers are stated on the type label of the instrument, inside the housing 

or on the delivery note.

Caution:

The order data and factory data contain important presettings for the 

sensor. Reliable operation and correct measurement are not possible 

without these data.

Electronics modules are adapted to the respective sensor. They differ, 

e.g. in the signal output, the power supply or approval.
First of all check by means of the overview in chapter "

Mounting prep-

arations

", if you are using the suitable electronics module. Compare 

the new electronics module with the existing one. The specifications 

on the type label must correspond. This applies mainly to instruments 

with approvals.

Warning:

Switch off voltage supply before starting the installation procedure. 

The replacement electronics may only be installed when the sensor is 

in a 

de-energised state

. Non-observance will damage the electron-

ics!

4.2  Mounting preparations VEGACAL

Electronics module CL-E.60H. is suitable for VEGACAL - 4 … 20mA/

HART.
Electronics module CL-E.60P.  is suitable for VEGACAL - Profibus PA.
Electronics module CL-E.60F. is suitable for VEGACAL - Foundation 

Fieldbus FF.
Electronics module CL-E.60X. suitable for VEGACAL - with signal 

conditioning instrument.

Sensor serial number

Assignment

Electronics versions

Содержание VEGACAL 60 Series

Страница 1: ...Operating Instructions Electronics module VEGACAL series 60 Document ID 30531...

Страница 2: ...iption 3 1 Configuration 5 3 2 Principle of operation 5 3 3 Packaging transport and storage 5 4 Mounting 4 1 General instructions 6 4 2 Mounting preparations VEGACAL 6 4 3 Installation procedure 7 5 S...

Страница 3: ...m 1 3 Symbols used Information tip note This symbol indicates helpful additional information Caution If this warning is ignored faults or malfunctions can result Warning If this warning is ignored inj...

Страница 4: ...3 Safety instructions for Ex areas Please note the Ex specific safety information for installation and op eration in Ex areas These safety instructions are part of the operating instructions manual a...

Страница 5: ...es on the transport packaging Nonobservance of these instructions can cause damage to the device The delivery must be checked for completeness and possible transit damage immediately at receipt Ascert...

Страница 6: ...or Reliable operation and correct measurement are not possible without these data Electronics modules are adapted to the respective sensor They differ e g in the signal output the power supply or appr...

Страница 7: ...ow the respective position of the electronics compartment in a single or double chamber housing 1 Fig 1 Single chamber housing 1 Position of the electronics module 1 Fig 2 Double chamber housing 1 Pos...

Страница 8: ...ning levers 6 Insert the new electronics module carefully 7 Screw in the two holding screws and tighten them 8 Connect the connection cables according to the operating in structions manual of the resp...

Страница 9: ...After entering the serial number the order data of the sensor are displayed The serial number is also available directly from VEGA Below the order data you will find Sensor data for Service DTM as XML...

Страница 10: ...doing this you help us carry out the repair quickly and without hav ing to call back for needed information If a repair is necessary please proceed as follows Print and fill out one form per instrumen...

Страница 11: ...rective 2002 96 EG This instrument is not subject to the WEEE directive 2002 96 EG and the respective national laws in Germany e g ElektroG Pass the instrument directly on to a specialised recycling c...

Страница 12: ...12 8 Supplement Electronics module VEGACAL series 60 30531 EN 170531 8 Supplement 8 1 Technical data Technical data are stated in the operating instructions manual of the respective sensor...

Страница 13: ...13 Notes Electronics module VEGACAL series 60 30531 EN 170531...

Страница 14: ...14 Notes Electronics module VEGACAL series 60 30531 EN 170531...

Страница 15: ...15 Notes Electronics module VEGACAL series 60 30531 EN 170531...

Страница 16: ...of delivery application practical use and operat ing conditions of the sensors and processing systems correspond to the information available at the time of printing Subject to change without prior n...

Отзывы: