50
8 Maintenance and fault rectification
VEGABAR 66 • 4 … 20 mA/HART - climate compensated
36738-EN-130321
F
perf
= √((F
T
)
2
+ (F
Kl
)
2
(+ F
a
)
2
)
With:
•
F
total
: Total deviation
•
F
perf
: Basic accuracy
•
F
stab
: Long-term drift
•
F
T
: Temperature coefficient (influence of medium or ambient
temperature)
•
F
Kl
: Deviation
•
F
a
: Error current output
Level measurement 1.5
mWs
Ambient temperature 40
°C, product temperature 10
°C
VEGABAR 66 with measuring range 0.2
bar
Deviation <
0.1
%
Calculation of the set Turn Down:
TD = 200
mbar/147
mbar, TD = 1.4
Basic accuracy digital output signal in percent:
F
perf
= √((F
T
)
2
+ (F
Kl
)
2
)
F
T
= (0.15 + 0.4 x TD)
%
F
Kl
= 0.1
%
F
perf
= √((0.15 % + 0.4 % x 1.4)
2
+ (0.1
%)
2
)
F
perf
= 0.72
%
Total deviation digital output signal in percent:
F
total
= F
perf
+ F
stab
F
stab
= (0.1
%
x
1.4)/year
F
stab
= 1.4
%
F
total
= 0.72
% +
1.4
% = 2.12
%
Total deviation digital output signal absolute:
F
total
= 2.12
%
x
147
mbar/100
% = 3.11
mbar = 31.77
mm
Basic accuracy analogue output signal in percent:
F
perf
= √((F
T
)
2
+ (F
Kl
)
2
+ (F
a
)
2
)
F
T
= (0.15 + 0.4 x TD)
%
F
Kl
= 0.1
%
F
a
= 0.15
%
F
perf
= √((0.15 % + 0.4 % x 1.4)
2
+ (0.1
%)
2
+ (0.15
%)
2
)
F
perf
= 0.73
%
Total deviation analogue output signal in percent:
F
total
= F
perf
+ F
stab
F
stab
= (0.1
%
x
1.4)/year
F
stab
= 1.4
%
F
total
= 0.73
% +
1.4
% = 2.13
%
Example