PLICSCOM
enables the adjustment without
fi
lling or pressure
.
Y
ou can carry out the settings in the workshop without the
instrument having to be installed
.
I
n addition to the menu items for zero
/
span adjustment
,
the
current measured value is displayed
.
T
o switch over to another adjustment unit
(
in the example from
bar to mbar
),
proceed as follows
:
5
)
1
P
ush the
[
OK
]
button in the measured value display
,
the
menu overview is displayed
.
▶
B
asic adjustment
D
isplay
D
iagnostics
S
ervice
I
nfo
2
C
on
fi
rm the menu
"
B
asic adjustment
"
with
[
OK
]
,
the
menu item
"
U
nits of measurement
"
will be displayed
.
U
nits of measurement
bar
3
A
ctivate the selection with
[
OK
]
and select the requested
unit with
[–
>
]
(
in the example mbar
).
4
C
on
fi
rm with
[
OK
]
and move to position correction with
[–
>
]
.
T
he adjustment unit is now changed from bar to mbar
.
P
roceed as follows
:
1
E
dit the o
ff
set value in the menu item
"
P
osition correction
"
with
[
OK
]
.
P
osition correction
P
O
ff
set
=
+
0000
.
0
mbar
53
.
0
mbar
2
S
et the requested o
ff
set value
,
e
.
g
.
the actual measured
value with reciprocal sign with
[–
>
]
and
[
+
]
.
3
C
on
fi
rm with
[
OK
]
and move to min
. (
zero
)
adjustment with
5
)
S
election options
:
mbar
,
bar
,
psi
,
P
a
,
k
P
a
,
MP
a
,
in
H
g
,
mm
H
g
,
in
H
2
O
,
mm
H
2
O
.
S
elect the adjustment unit
C
arry out position correction
24
VEGABAR
53
-
P
ro
fi
bus
PA
S
etup with the indicating and adjustment module
PLICSCOM
28250
-
EN
-
051004
Содержание VEGABAR 53
Страница 1: ...Operating Instructions VEGABAR 53 Pro bus PA p...
Страница 57: ...Manufacturer declaration Fig 27 Manufacturer declaration VEGABAR 53 Pro bus PA 57 Supplement 28250 EN 051004...
Страница 58: ...10 5 Industrial property rights 58 VEGABAR 53 Pro bus PA Supplement 28250 EN 051004...
Страница 59: ...VEGABAR 53 Pro bus PA 59 Supplement 28250 EN 051004...