41
13 Supplement
VEGABAR 29 • Two-wire 4 … 20 mA
57087-EN-210301
Switch-on phase
Start-up time with operating voltage U
B
≤ 2 s
Starting current for run-up time
≤ 3.6 mA
Output variable - two-wire 4 … 20 mA
Output signal
4 … 20 mA - passive
Connection technology
Two-wire
Range of the output signal
3.8 … 20.5 mA
Signal resolution
5 µA
Fault signal, current output (adjustable)
≤ 3.6 mA, ≥ 21 mA, last measured value
Max. output current
21.5 mA
Load
See load resistance under Power supply
Transferred measured value
Pressure
Dynamic behaviour output
Dynamic characteristics - Current output
6)
90 %
100 %
10 %
t
t
T
A
t
t
S
2
1
Fig. 26: Behaviour the current output in case of sudden change of the process variable. t
T
: dead time; t
A
: rise time;
t
S
: jump response time
1 Process variable
2 Output signal
Size
Time
Dead time
≤ 2 ms
Rise time (10 … 90 %)
≤ 4 ms
Step response time (ti: 0 s, 10 … 90 %)
≤ 6 ms
Damping (63 % of the input variable)
0 … 9 s, adjustable
Reference conditions and influencing variables (according to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1
Ʋ
Temperature
+18 … +30 °C (+64 … +86 °F)
6)
Depending on medium and temperature