6
2 Mounting
Mounting adapter PTB 9 • For mounting on the existing mounting bracket (GM-16™)
55642-EN-170815
1
3
2
4
4
5
12
2
10
13
7
14
11
10
9
8
7
6
Fig. 4: Mounting of the mounting adapter with anti-vibration clamp
1 Screw M6 x 160 (2 pieces)
or screw ¼" x 6½" (2 pieces)
2 Lock washer - Nordlock for M6 or ¼" (4 pieces)
3 Cover plate (1 piece)
4 Clamp collars (2 pieces)
5 Base (1 piece)
6 Mounting adapter PTB 9
7 Lock washer - Nordlock for M8 or 5/16" (4 pieces)
8 Screws M8 x 30 (2 pieces)
or screw 5/16" x 1¼" (2 pieces)
9 Mounting bracket for GM-16 (provided by customer)
10 Washer - provided by customer (recommendation: Nordlock M12 or ½",
4 pieces)
11 Hexagon nut - provided by customer (recommendation: M12 or ½",
2 pieces)
12 Hexagon nut M6 or 5/16" (2 pieces)
13 Screw - provided by customer (2 pieces)
Recommendation: M12 x 30 or ½" x 1¼"
14 Hexagon nut M8 or 5/16" (2 pieces)
Danger:
Before mounting, make sure that the corresponding radiation source
is securely closed. Use a padlock to secure the source holder in the
closed condition and prevent it from being inadvertently opened. Also
take note of the instructions in the operating instructions manual of
the source holder.
1. Place the mounting adapter (6) according to the above illustration
onto the existing mounting bracket (9).