Surge
IEC 61000-4-5
± 1 kV line(s)
to
line(s)
± 2 kV line(s)
to earth
Not applicable
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage
dips,
short
interruptions
and
voltage
variations
on
power
supply
input lines
IEC 61000-4-11
<5 % UT
(>95 % dip in
UT)
for 0.5 cycle
40 % UT
(60 % dip in
UT)
for 5 cycles
70 % UT
(30 % dip in
UT)
for 25 cycles
<5 % UT
(>95 % dip in
UT)
for 5 s
Not applicable
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the device requires continued
operation during power mains
interruptions, it is recommended
that the device be powered from
a continuous power supply or a
battery.
Power
frequency
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial
or
hospital