Disinfections: 

 

Please follow the following steps to disinfect your CPAP unless otherwise 
specified by your physician. It is suggested that the unit is disinfected after the last 
treatment of the day. 
1.  Use one part white vinegar with three parts distilled water. Make sure mixed 

solution is enough to submerge the air tube, mouthpiece and mask. 

2.  Complete Rinsing Steps 1-3. 
3.  Wash air tube, mouthpiece and mask in warm water and a mild detergent. 

Then wash them in hot tap water. 

4.  Submerge these parts in the vinegar and water solution for thirty minutes. 
5.  Complete Rinsing Steps 3-5. 

 

 

Filter change   

 

1.  Do not wash or clean the filter. And do not operate without a filter. 
2.  Change the filter every 30 days or when the filter turns gray. 
3. Changing procedure: 

A. Replace the used filter with a new one. 

 

VII REPLACEABLE PARTS & ACCESSORIES

 

 
In order to provide appropriate treatment, please use replaceable parts and 
accessories recommended below: 

 

Air tube: 70253- Breathing/Anesthesia Tubing,    HSINER CO., LTD 
 

   Mask:  CPAP/VPAP Masks(FDA 510(K) No. K063268), HSINER CO., LTD 
 

Power Adapter: UE24WCP-120167SPA, Dongguan Shilong Fuhua 

Electronic Co.,Ltd. 

 
 
 
 
 
 
 
 
 

   Heated Humidifier:  HH-02 Humidifier,   Vega Technologies Inc 

Содержание CP-03

Страница 1: ...1 INSTRUCTION VEGA MANUAL Continuous Positive Airway Pressure MODEL CP 03 READ THIS INSTRUCTION Version 1 0 Date Nov 04 2009 ISO9001 ISO13485 Certified This device complies with EU EMC standard EN6060...

Страница 2: ...device uses a motor which drives a fan to create a continuous air pressure to force the air into the patient s lung through the nasal cavity so that the patient s airway is expedite which is the key t...

Страница 3: ...R404 1 B Assemblies...

Страница 4: ...cts are being used 4 Disconnect the unit from the electrical outlet before cleaning filling and after each use 5 Do not use other attachments unless recommended by manufacturer Operating Cautions 1 Co...

Страница 5: ...upply 12V DC 2 Current Consuming Max 1 5A 3 Air Pressure Range 3 cmH2O 20 cmH2O 4 Ramp Time 0 60 min 5 Air Pressure Calibrate 3 cmH2O 6 Working Flow 3 0 3m 3 h 9 cmH2O 7 Noise 33 dB 9 cmH2O 1m away 8...

Страница 6: ...nd the device returns to the standby mode 4 Press RAMP at the normal working mode CPAP runs at the ramp mode the LCD outputs RAMP and Pressure xx x cmH2o xx x is the numeric value of the air pressure...

Страница 7: ...ing on the screen the LCD displays Mask Off meanwhile the backlight flashes at the frequency of 1 Hz for 1 minute The following are configuration operations 5 Press RAMP and at the same time in standb...

Страница 8: ...feet 2500 5000 feet 5000 7500 feet and 7500 9000 feet respectively Set this with and according to the altitude of your current location 8 Press MODE at the altitude setting mode the device enters the...

Страница 9: ...Press MODE at the ramp time setting mode the device enters the starting air pressure setting mode The LCD displays Ramp Start Pressure xx x cmH2o xx x is the numeric value of the starting air pressur...

Страница 10: ...he operation time checking mode The LCD displays Operation Hrs xxxx x xxxx x is the operation time of the device This value is not to be changed but only to be reset Press for a period of time of 2 se...

Страница 11: ...on the screen Disconnect the USB cable the device goes back to Standby mode VI Cleaning and Sterilization It is recommended that the air tube and mask be thoroughly cleaned with hot water after each u...

Страница 12: ...Submerge these parts in the vinegar and water solution for thirty minutes 5 Complete Rinsing Steps 3 5 Filter change 1 Do not wash or clean the filter And do not operate without a filter 2 Change the...

Страница 13: ...R404 1 Consult accompanying documents Type BF equipment Disposal for separate collection Device in protection class 2...

Отзывы: