R404-1
sealed container.
Disinfections:
Please follow the following steps to disinfect your CPAP unless otherwise
specified by your physician. It is suggested that the unit is disinfected after the last
treatment of the day.
1. Use one part white vinegar with three parts distilled water. Make sure mixed
solution is enough to submerge the air tube, mouthpiece and mask.
2. Complete Rinsing Steps 1-3.
3. Wash air tube, mouthpiece and mask in warm water and a mild detergent.
Then wash them in hot tap water.
4. Submerge these parts in the vinegar and water solution for thirty minutes.
5. Complete Rinsing Steps 3-5.
Filter change
1. Do not wash or clean the filter. And do not operate without a filter.
2. Change the filter every 30 days or when the filter turns gray.
3. Changing procedure:
A. Replace the used filter with a new one.
VII REPLACEABLE PARTS & ACCESSORIES
In order to provide appropriate treatment, please use replaceable parts and
accessories recommended below:
Air tube: 70253- Breathing/Anesthesia Tubing, HSINER CO., LTD
Mask: CPAP/VPAP Masks(FDA 510(K) No. K063268), HSINER CO., LTD
Power Adapter: UE24WCP-120167SPA, Dongguan Shilong Fuhua
Electronic Co.,Ltd.
Heated Humidifier: HH-02 Humidifier, Vega Technologies Inc