FX87 DWDM Tunable Laser Source User’s Manual
Page
20
of
39
•
REF +/-
command: adjusts the reference level finely by 0.01 increments with the help of
the
|Up|
and
|Down|
buttons.
8.3.2 Saving Measurement Results
To save the result, press the
|Shift/Enter|
and
|Up|
buttons simultaneously. The result is then
saved with the comment (note) from the previously saved result.
8.4 Using WaveID
8.4.1 FX87 Device and FX-8x OPM Connected from End to End
FX87 connected over fiber under test to FX-8x OPM
Set the FX-8x OPM
WAVE
|F1|
function to
AUTO
, and choose from one of two FX87 transmit
setups
1. Set the FX87 in
DWDM OLS
mode to transmit by pressing
LAS ON
|F1|
, and press
CHANNEL
|F2|,
and
navigate to the desired DWDM channel using the
|Up|
and
|Down|
keys and press
|Shift/Enter|
to confirm the selection
.
2. Set the FX87 in
PM1/PM2
(
OPM
) mode and press
LASER
|F2|
and choose either
LAS ON
|F1|
to turn on the laser for the current displayed wavelength by the
>
…. <
or navigate to the desired DWDM channel using the
|Up|
and
|Down|
keys and press
|Shift/Enter|
to confirm the selection
.