Installation
Connecting the Dimension 3100 System
Rev. D
Dimension 3100 Manual
67
2. Toggle the vacuum power switch, located on the front of the microscope unit, to
OFF
(see
Figure 4.4g
Vacuum Power Switch
Route the VT103 Air Table Cabling
Figure 4.4h
Cable Clamp
1. Route all cables connected to the Dimension 3100 microscope through the slot located in the
far right region of the VT103 vibration isolation table.
2. Position the Dimension 3100 microscope in the desired location.
3. Mark the cables with a marker to reference each cable’s position in the cable clamp (see
Note:
When secured in place, the cable clamp is located underneath the vibration
isolation table, near the aforementioned slot (see
3918
VACUUM
ON
OFF
Vacuum Power Switch
3907
NanoScope IIIa Controller to
Dimension 3100 Microscope
Cable, 37-pin D
Remaining Dimension 3100 Cables
Содержание Dimension 3100