Contact AFM
Basic Contact Mode AFM Operation
Rev. D
Dimension 3100 Manual
123
8.1.9 Set Initial Scan Parameters
Scan Controls Panel
In the
Scan Controls
panel, set the following initial scan parameters (see
1. Set the
Scan Rate
to
2 Hz
.
2. Set the
Scan Size
to
1µm
.
3. Set the
Scan Angle
to
0
.
4. Set
X
and
Y Offsets
to
0
.
Figure 8.1c
Suggested Scan Controls Settings
Other Controls Panel
Keep the
Z limit
at its maximum value.
Figure 8.1d
Suggested Other Controls Settings
017
Содержание Dimension 3100
Страница 12: ...xii Dimension 3100 Manual Rev D ...
Страница 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Страница 72: ......
Страница 106: ......
Страница 118: ......
Страница 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Страница 222: ......
Страница 266: ......
Страница 274: ......