Scan Routine Parameters
7-15
NOTE –
The
End Force
that will be applied by the stylus is calculated and displayed.
Figure 7-16: Scan Parameter Dialog Box: Deflection Parameters (Optional)
Example:
•
Scan Duration
is set at 13 seconds.
•
Stylus Force
is set at 3 mg (on the
Nominal Parameters
tab).
•
Interval
is set at 1000 msec (1 second).
•
Increment
is set at 0.05 mg.
•
End Force
is calculated and displayed as 3.65 mg (13 seconds divided by 1 second, multiplied
by 0.05 mg = 0.65 mg to be added in 0.05-mg increments to 3 mg).
DISPLAY PARAMETERS
In its
Display Parameters
section, the
Scan Routines
window contains parameters that allow
automatic manipulation of the graphic display of the profile trace. The display parameters are
described below.
Содержание Dektak 150
Страница 1: ......
Страница 2: ...P N 980 294 Standard P N 980 298 Cleanroom Revision A DEKTAK 150 SURFACE PROFILER USER S MANUAL ...
Страница 7: ......
Страница 17: ......
Страница 81: ......
Страница 105: ......
Страница 149: ......
Страница 191: ......
Страница 251: ......
Страница 257: ...A 6 Facilities Specifications Figure A 6 Dektak 150 without Enclosure Side View ...
Страница 258: ...Facilities Specifications A 7 Figure A 7 Dektak 150 Dimensions without Enclosure Top View ...
Страница 259: ...A 8 Facilities Specifications Figure A 8 Dektak 150 without Enclosure Center of Mass ...
Страница 273: ......
Страница 283: ......
Страница 320: ...3D Mapping Function E 37 Figure E 26 Database Options Button Figure E 27 Database Options Dialog Box ...
Страница 321: ......
Страница 331: ......
Страница 333: ......
Страница 336: ......
Страница 337: ......
Страница 349: ......