Head, Probe and Sample Preparation
Initial Preparation for Contact AFM Imaging
74
MultiMode SPM Instruction Manual
Rev. B
Figure 5.1c
Head is Held Securely Using Retaining Springs
Check Head for Free Vertical Movement
Verify basic function of the motorized Z-axis by toggling the
Up
switch on the MultiMode base
(see
). This activates the leadscrew at the rear of the unit to lift the head upward. If you
are using a standard, three-screw scanner, the two forward screws will also have to be rotated to
keep the head level while lifting. (Single-screw, vertical scanners require only that the rear,
motorized screw be rotated to raise and lower the head.) To verify rotation of the motorized screw,
feel the
fl
exible coupling on the base with your
fi
nger while you toggle the
Tip
Up
/
Down
switch.
Figure 5.1d
Tip Up / Down Switch on the MultiMode SPM’s Base
Reattach retaining
springs (2)
Tip up
Tip down