OFF – the port is turned off; 
Pr1 – Protocol VEDIA with uninterrupted weight transmission; 
Prn – is not used. 

After each change of the protocol, the scale is restarted. 

 - In the submenu, an operating mode for the backlights of the two LCD displays is 

selected. Available working modes: 
1.OFF – Backlight is turned off each time the readings for the weight is zero (battery saving mode); 
2.On – Backlight is always on. 

 - In the submenu the user has access to the following service mode functions: 

1. ADC – Test of the Analog-to-Digital Converter. The display shows direct readings of the 
ADC; 
2. t

o

C – Thermometer. The display shows themperature in degree Celsius; 

3. Ver – Version. The display shows the current version of the scale software; 
4. nCal – The display shows the number of the performed actions in service mode. 

 

 

 

6. Operation with the rechargeable battery 
 

      The charging of the rechargeable battery is made with an adapter, bought together with the scale. 
The adapter should be plug in DC 12V jack that is on the right side of the case. During the recharging 
the scale can operate or can be switched off. The battery is fully recharged for approximately 8 hours. 
The current voltage of the battery can be checked from menu 1.Bat of the scale. When “battery” 
symbol is shown on the display, please recharge the battery as soon as possible. In reaching 
impermissibly low voltages, the display turns off and the scale operation is blocked till the normal 
values of the voltage are restored (plug in the adaptor). 
 
 

 

7. Errors 
 

 

Err1 – it is shown after passing the indication test when the load cell is deformed of 
overloading or due to some reason the platform is heavier. The extra load should be removed. 
If the error does not disappear turn to the closest service centre.  

 

Err2 – when it is shown after passing the indication test it means that the load cell is 
deformed backwards or the platform has is lightened. Place the pan on the scale or if it is on 
its place turn to the closest service centre. 

 

 Err3 – after the initial zero- setting the weight of the platform is below the given limits. 
Probably the pan has been removed during the operation or something is raising it. 

 

Err4 – the scale is loaded with so extremely heavy load that it cannot be read by the 
electronic block. If the scale is unloaded, turn to the nearest service. 

 

Err5 – It is shown on the display with turning on the scale when data incorrectness in the 
memory is detected.  

 

“-HI-“ – overloading indication –a load over the maximum capacity is placed on the platform 

 

 
 
 
 
 
 
 
 
 

Содержание VDI 3/6

Страница 1: ...User Manual ELECTRONIC SCALE The product is designed and produced in the terms of applied Quality Management System ISO 9001 2008...

Страница 2: ...ion of conformity In case of missing please immediately turn to the distributor you bought the scale from or directly to VEDICOM Ltd II Brief Introduction Electronic scales VEDIA are non automatic wei...

Страница 3: ...ion Platform size mm No of Load Cells Built in battery VDI 3 6 3 6kg 1g 2g 350 450 1 yes VDI 6 15 6 15kg 2g 5g 350 450 1 yes VDI 6 30 6 30kg 2g 10g 350 450 1 yes VDI 15 30 15 30kg 5g 10g 350 450 1 yes...

Страница 4: ...IV Layout of the scale and its components Keyboard Display...

Страница 5: ...precise instrument Please place it on flat surface and avoid serious hits Loads over the maximum capacity can damage the load cell permanently Avoid operating it on vibrating surfaces and close to sou...

Страница 6: ...s the measured weight The weight data are considered correct when the indicator for stable reading lights up 5 Work with the menus and service mode functions of the scale To enter the main menu of the...

Страница 7: ...rrent voltage of the battery can be checked from menu 1 Bat of the scale When battery symbol is shown on the display please recharge the battery as soon as possible In reaching impermissibly low volta...

Страница 8: ...LGARIA 1784 Sofia 131 Tzrigradsko shose centre AD 5th floor office 526 tel fax 02 965 85 57 58 5800 Pleven 21 San Stefano Street office 5 Tel fax 064 846080 064 845445 e mail office vedicom org http w...

Отзывы: