zeroing successfully.
* Grounding first and then press “ZERO” key about two
seconds and avoid objects with electricity to zero. After
un-pressing the key, the meter sounds and displays 0 if
zeroing successfully.
* Connect the conductor to the “Ground of the meter” first
and then press “ZERO” key about two seconds and test the
conductor to zero. After un-pressing the key, the meter
sounds and displays 0 if zeroing successfully.
For example: one hand handholds the meter and put the
other hand as conductor in front of the meter to zero.
3 The zeroed method of the ion balance testing:
Insert the ion balance testing plate firstly and then connect the
“front testing plate” to the “Ground of the meter”. Press
“ZERO” key about two seconds to zero. After un-pressing the
key, the meter sounds and displays 0 if zeroing successfully.
4.3 switch between the static testing and the
ion balance testing
1 the meter runs into the static testing after switching on the
power supply.
2 press “
STATIC/IB
” key to the ion balance testing (display
“IB”).
3 press “
STATIC/IB
” key again to return to the static testing.
4 press “
STATIC/IB
” key again to the ion balance testing.
4.4 HOLD/MAX
HOLD:
keep the current testing value for recording it. And
lock the functions of “STATIC /IB” and “ZERO” until
exiting the HOLD process.
ENTER:
press “HOLD/MAX” key (less than 1.5s) to
hold process (display “HOLD” after
un-press the key).
EXIT:
press “HOLD/MAX” key to exit the hold
process (without display “HOLD”).
press “
STATIC/IB
” key
Static testing
Ion balance testing