Pax-i3D Smart(PHT-30LFO) Installation Manual
64
4 Installing the Equipment: Floor Standing (Optional)
4.7
Leveling the Equipment
8mm / 0.3"
T-shaped hex wrench
Spirit level
Ensure that the spirit level should restonly on the locations indicated in
the following figures to obtain the accurate center.
1.
Prepare the spirit level.
Tube head
2.
Position the rotating unit so that the X-Ray
tube head faces the front.
3.
Turn all eight screws on the base plate unit
clockwise until they touch the ground.