PaX-i
(PCH-2500) Installation Manual
181
Appendix
B. Installing the Emergency Switch
●
Install the emergency switch stop switch on the power cable line.
●
Install this switch so that it is easy to reach in the emergency case but can’t be pushed by
mistake.
●
The switch shall be a type of mistake-proof.
●
The switch is not supplied.
●
The switch shall be installed at a height of 1.2 to 1.5 meters(47 to 60”).
To Pax-I
Plug
Receptacle
L
L
N
N
PE
PE
Second
Protective ground wire
Centra distribution panel
w/ 25A circuit breaker
Emergency shutdown
1. The cable sizes: N, L and PE≥ 12 AWG(3 x 4 mm
2
).
2. The cable to the emergency switch shall be the same size as the power cable itself.
3. Install the socket connector terminal for the 2nd protective ground wire.
Содержание Pax-i PCH-2500
Страница 1: ...Installation Manual Full version Model PCH 2500 Version 3 06 English...
Страница 2: ......
Страница 3: ......
Страница 46: ...This page is intentionally left blank...
Страница 82: ...This page is intentionally left blank...
Страница 104: ...This page is intentionally left blank...
Страница 124: ...This page is intentionally left blank...
Страница 166: ...This page is intentionally left blank...
Страница 167: ...Acquiring the Test Image 10...
Страница 169: ...11 Technical Specifications...
Страница 176: ...This page is intentionally left blank...
Страница 179: ...PaX i PCH 2500 Installation Manual 179 Appendix Schematic diagram Components supplied...
Страница 183: ...PaX i PCH 2500 Installation Manual 183 Appendix 2 3M 2 2M 2 1M 2 0M 4 Separate the side cover...
Страница 199: ......
Страница 200: ...Postal code 18449 13 Samsung 1 ro 1 gil Hwaseong si Gyeonggi do Korea www vatech co kr...