A. Installing the Warning Lamp and Door Interlock Switch
Requirements:
1. The warning control system shall be connected to the ERB (earth reference bar) of the room
that it is associated with.
2. The switching arrangements, location, height, and a number of illuminated warning signs shall
be agreed with the local radiation protection advisor (RPA). (customer)
3. A fluorescent lamp shall not be used in the ‘X-rays on’ sign.
4. The customer shall be responsible for the proper installations for the warning control system,
including the lamp and door interlock switch, based on the MEIGaN guideline.
MEIGaN: Medical Electrical Installation Guidance Notes
5. Pre-installation planning is crucial to the successful installation of these devices.
6. For further details, refer to the accompanying volume: Specification for Electrical Installation
Block diagram:
Содержание Pax-i PCH-2500
Страница 1: ...Installation Manual Full version Model PCH 2500 Version 3 06 English...
Страница 2: ......
Страница 3: ......
Страница 46: ...This page is intentionally left blank...
Страница 82: ...This page is intentionally left blank...
Страница 104: ...This page is intentionally left blank...
Страница 124: ...This page is intentionally left blank...
Страница 166: ...This page is intentionally left blank...
Страница 167: ...Acquiring the Test Image 10...
Страница 169: ...11 Technical Specifications...
Страница 176: ...This page is intentionally left blank...
Страница 179: ...PaX i PCH 2500 Installation Manual 179 Appendix Schematic diagram Components supplied...
Страница 183: ...PaX i PCH 2500 Installation Manual 183 Appendix 2 3M 2 2M 2 1M 2 0M 4 Separate the side cover...
Страница 199: ......
Страница 200: ...Postal code 18449 13 Samsung 1 ro 1 gil Hwaseong si Gyeonggi do Korea www vatech co kr...