background image

 

VATECH Co., Ltd.

                                                                                                                         

Release 

V2.0.0

 

PaX-500 

3. Cautions 

 

Make sure to observe following. 

 

 Make sure to follow the specified process of operation for safety of users and patients.   

 

 Make sure to check the status of products such as power, PC and cable before use.   

 

 

Perform the operation while it is in stop status. Violation of this may result in 

malfunction.  

 

 Make sure to proceed to the next imaging status after passage of Cooling Time (process of 

coolding down X-Ray tube) according to voice announcement after imaging.   

 

 Place this product away from water, moisture or foreign substance since this product is a 

precision medical electronic device. 

 

 

 Immediately cut off the power if the product is exposed to water or foreign substance during 

use, resulting in abnormal operation. Contact the agent for support.   

 

Содержание PaX-500

Страница 1: ......

Страница 2: ......

Страница 3: ...ht or Packaging Unit Indicates in product 5 Performance and Directions of Use Indicates in user s manual and catalog 6 Cautions for Use Indicate in user s manual 7 Other Necessary Information 1 Rated...

Страница 4: ...e use for the safety of user and patient 2 Make sure to comply with cautions and procedure to use indicated in this manual and the product 3 Do not use the product while it abnormally operates or is i...

Страница 5: ...e sure to become familiar with cautions and procedure of use Also note that it may slightly differ from the contents of this manual depending on specification of product Following marks are used for e...

Страница 6: ...n 25 1 Preparation before Imaging 26 2 How to Align Patients 27 2 1 How to Align Patients on Panoramic unit 27 2 2 How to Align Laser Beam 29 2 3 How to Align Patients on Cephalometric unit 31 3 How t...

Страница 7: ...User manual 5 1 Chapter 1 Product Overview This Chapter describes overall features components of products as well as various cautions in details...

Страница 8: ...inition sensor developed independently providing wider length and width than film without the jaw part being cut Support more accurate diagnosis imaging than LCD as well as voice announcement function...

Страница 9: ...ication of this product may be subject to change without any notice for improvement of product performance Software Components Viewer program EasyDent V4 program Imaging program EasyDent V4 imaging pr...

Страница 10: ...atus Violation of this may result in malfunction Make sure to proceed to the next imaging status after passage of Cooling Time process of coolding down X Ray tube according to voice announcement after...

Страница 11: ...rocess Children or pregnant women should consult with the doctor in charge before imaging A worker should be at least 2 m 6 feet away from the equipment during imaging Equipment should be located insi...

Страница 12: ...ing cases If the product is installed by our authorized agent If it is installed in accordance with cautions and conditions for installation If genuine product is used as approved by our company If ma...

Страница 13: ...User manual 11 2 Chapter 2 Components and Functions This Chapter describes hardware components and software architecture of the product in details...

Страница 14: ...ric image LED Lamp LED for checking X Ray irradication Headset part for fixing the patient head during cephalometric imaging Chinrest part for fixing the patient head during panoramic imaging LCD pane...

Страница 15: ...ow Show imaging progress status for capturing image and show the image captured after imaging show image captured real time at the time of actual imaging Pateint Info Window Show information of patein...

Страница 16: ...ection Special mode Maxillary clear Canal clear Incisor clear Orthogonal are available for selection Select Arch Button Normal Wide Narrow Child are available for selection Lamp Button Turn the lamp O...

Страница 17: ...User manual 15 This software architecture may differ depending on specification of your product and be subject to change without notice for improvement of product performance...

Страница 18: ...ptured after imaging show image captured real time at the time of actual imaging Pateint Info Window Show information of pateint such as name age and chart of pateint to be taken Imaging Instruction W...

Страница 19: ...rangement of pateints Select Imaging Mode Button Select one of imaging mode such as Panorama or Cephalo Confirm Button Confirm the current setting Ready Button Perform imaging according to instruction...

Страница 20: ...value suitable for patient etc Mode button Select the imaging mode Pano Standard Pano Special Cephalo Image button Show the image currently saved in the memory can save up to 10 images Return button...

Страница 21: ...s Pano Special Select the special imaging mode Orthogonal Incisor Clear Maxillary Clear Canal Clear Cephalo Select the Cephalo imaging mode Lateral Frontal SMV Carpus Next Go to the next stage Select...

Страница 22: ...r joint TMJ of patient Show the images in the order of Right Open Right Close Left Close Left Open in one sheet of image For imaging with LCD panel it displays Open image only at the time of TMJ Open...

Страница 23: ...User manual 21 Next Go to the next stage Select Next to go to the next stage Before Go to the previous stage Select Before to cancel the current setting value...

Страница 24: ...ap Incisor clear The imaging mode of optimizing the foreteeth of the pateint Maxillary clear The imaging mode of optimizing maxillary sinus of the pateint Canal clear The imaging mode of optimizing ca...

Страница 25: ...ontal image of the head at the time of measuring pateint s head SMV The mode of taking the caudocranial image at the time of measuring pateint s head Carpus The imaging mode of checking the growth of...

Страница 26: ...creen X ray Setting Select kVp and mA of X ray manually Patient Setting Select the feature gender bone density of the patient Confirm Confirm the current setting and go to the next stage Exit Cancel t...

Страница 27: ...Chapter 3 Basic Operation This Chapter describes basic operation of the product as well as important functions for easy understanding and use of product by users Refer to this Chapter for effective us...

Страница 28: ...is turned on Do not perform any operation of PC while the product is in operation Check whether there is any problem in operation of PC When the equipment is already turned on select the Imaging Mode...

Страница 29: ...ight position It is better if the patient comes closer to the system as much as possible Have the patient hold the handle The patient may slowly fall down backward when he she does not hold the handle...

Страница 30: ...accurately located See the following figure Adjust the Canine Laser Beam so that it is accurately located at the canine part of the patient 1 Panoraic position the Canine Laser beam at the canine 2 S...

Страница 31: ...User manual 29 2 2 How to Align Laser Beam Panoramic Sinus Panoramic Sinus Positioning Jig Vertical Beam Horizontal Beam Canine Beam...

Страница 32: ...VATECH Co Ltd Release V2 0 0 PaX 500 30 TMJ open TMJ close TMJ open TMJ close Positioning Jig Vertical Beam Horizontal Beam Canine Beam...

Страница 33: ...c Imaginary Horizontal Baseline The highest position of earhole and a boundary region between cheek and proturant part of an eye Effect of improving uniformity radiograph since horizontal anatomical m...

Страница 34: ...VATECH Co Ltd Release V2 0 0 PaX 500 32 Frontal mode on Cephalometric unit Carpus mode on Cephalometric unit...

Страница 35: ...ients Follow the steps shown below to register new patient Turn on the product and computer Execute EasyDent V4 and click Patient icon to register new patient Enter the patient information when follow...

Страница 36: ...r specified Turn on the power switch of PaX 500C Leave as it is if power switch is on Execute EasyDent V4 program in PC Program is displayed as shown in the following figure Fig 1 Search and select th...

Страница 37: ...User manual 35 PC is checking system PaX 500 whether it is prepared to capture images or not Choose Panorama to capture panoramic images...

Страница 38: ...VATECH Co Ltd Release V2 0 0 PaX 500 36 After choosing Panorama system will change the imaging mode to Panoramic mode...

Страница 39: ...User manual 37 Set a condition of image capturing And then click Confirm The settings of X ray Gender and Bone density panoramic capturing mode and Arch Standard mode Special mode...

Страница 40: ...VATECH Co Ltd Release V2 0 0 PaX 500 38 System will calibrate a sensor for optimal condition of image capturing Position a patient on system...

Страница 41: ...eady button when the position of patient is finished Push an exposure switch to capture image after finishing the pareparation of system You should hold the exposure switch to capture image until the...

Страница 42: ...VATECH Co Ltd Release V2 0 0 PaX 500 40...

Страница 43: ...manual 41 A captured image will be shown on program after optimizing And click OK if you want to save the captured image to DB Click OK if you want to save the captured image to DB otherwise click Ca...

Страница 44: ...VATECH Co Ltd Release V2 0 0 PaX 500 42 After saving the program will ask you to capture image again If you want to capture again click OK Otherwise click Cancel...

Страница 45: ...ient in patient list after imaging the image list is reconfigured Images recently captured appear in the left as shown in the figure Double click on it to check the image in details Final Panoramic im...

Страница 46: ...Click Pano Ceph icon on the top left to execute imaging program For unregistered patient register him her in reference to user s manual of program separately prepared before capturing image Sensor mod...

Страница 47: ...User manual 45 PC is checking system PaX 500 whether it is prepared to capture images or not Choose Cephlao to capture cephalometric images...

Страница 48: ...VATECH Co Ltd Release V2 0 0 PaX 500 46 After choosing Cephalo system will change the imaging mode to Cephalometric mode...

Страница 49: ...l 47 Set a condition of image capturing And then click Confirn The settings of X ray Gender and Bone density cephalometric capturing mode System will calibrate a sensor for optimal condition of image...

Страница 50: ...VATECH Co Ltd Release V2 0 0 PaX 500 48 Position a patient on system Click Ready button when the position of patient is finished...

Страница 51: ...User manual 49 Push an exposure switch to capture image after finishing the pareparation of system You should hold the exposure switch to capture image till capturing of image...

Страница 52: ...VATECH Co Ltd Release V2 0 0 PaX 500 50...

Страница 53: ...be shown on program after optimizing And then click OK if you want to save the captured image to DB Click OK if you want to save the captured image to DB Otherwise click Cancel After saving the progra...

Страница 54: ...VATECH Co Ltd Release V2 0 0 PaX 500 52 If you want to capture again click OK otherwise click Cancel...

Страница 55: ...patient in patient list after imaging the image list is reconfigured Images recently captured appear in the left as shown in the figure Double click it to check the image in details Final Cephalometri...

Страница 56: ...creen Normal Capture the image of normal panorama Press the Button of image to capture if you want to capture other image Next Move to the next stage High Resolution Fast Scan Select this button to ca...

Страница 57: ...User manual 55 Confirm Select the contents of setting and move to the main screen...

Страница 58: ...re the Maxillary Clear image press the Button of image to capture if you want to capture other image Next Move to the next stage Maxillary Clear Right Left select the direction of imaging right or lef...

Страница 59: ...e to the next stage Lateral Select to capture the lateral image of head press the Button of image to capture if you want to capture other image Next Move to the next stage Select Pateint Select the X...

Страница 60: ...VATECH Co Ltd Release V2 0 0 PaX 500 58 4 Chapter 4 Emergency Measure This Chapter describes problems that may occur when using the equipment as well as their solutions...

Страница 61: ...s of Serial Port RS232 connecting PC and device If exposure switch is not working Cause Resolution Ready status Check whether it is ready for capturing at imaging program If imaging is not working Cau...

Страница 62: ...re Hard Normal Soft Man 72 kVp 8 mA 70 kVp 8 mA 68 kVp 8 mA Woman 70 kVp 8 mA 68 kVp 8 mA 66 kVp 8 mA Child 66 kVp 7 mA 66 kVp 7 mA 66 kVp 7 mA 1 3 Normal Ceph Gender Figure Hard Normal Soft Man 78 kV...

Страница 63: ...d written permission of the manufacturer and authors of this manual If you do not properly set the device causing it to malfunction or fail we cannot assume any responsibility VATECH Tel 82 31 679 207...

Страница 64: ......

Страница 65: ......

Страница 66: ......

Отзывы: