15.1 Recommended X-ray Exposure Tables
15.1.1 PANO Mode
Exposure Condition
PANO Option > Normal
Mode
Image
Option
Gender
/ Age
group
X-ray
Intensity
Tube Voltage
(kVp)
Tube Current
(mA)
PANO
Examination
High
Resolution
Man
Hard
75
12
Normal
74
12
Soft
73
12
Woman
Hard
74
12
Normal
73
12
Soft
72
12
Child
Hard
68
10
Normal
67
10
Soft
66
10
Green
(Optional)
Man
Hard
75
14
Normal
74
14
Soft
73
14
Woman
Hard
74
14
Normal
73
14
Soft
72
14
Child
Hard
68
12
Normal
67
12
Soft
66
12
Содержание GREEN SMART
Страница 1: ...User Manual Version 1 23 English...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 52: ...5 Imaging Software Overview 44 Green Smart User Manual Left blank intentionally...
Страница 90: ...7 Acquiring PANO Images 82 Green Smart User Manual Left blank intentionally...
Страница 110: ...8 Acquiring CEPH Images Optional 102 Green Smart User Manual Left blank intentionally...
Страница 126: ...9 Acquiring CBCT Images 118 Green Smart User Manual Left blank intentionally...
Страница 134: ...10 Acquiring 3D MODEL Scan Images 126 Green Smart User Manual Left blank intentionally...
Страница 142: ...12 Cleaning and Maintenance 134 Green Smart User Manual Left blank intentionally...
Страница 144: ...13 Disposing of the Equipment 136 Green Smart User Manual Left blank intentionally...
Страница 149: ...14 Technical Specifications Green Smart User Manual 141 ENGLISH Maximum Rating Charts Emission Filament Characteristics...
Страница 154: ...14 Technical Specifications 146 Green Smart User Manual Left blank intentionally...
Страница 185: ...15 Appendices Green Smart User Manual 177 ENGLISH...
Страница 198: ...15 Appendices 190 Green Smart User Manual STL Stereo Lithography TMJ Temporomandibular Joint UHD Ultra High Definition...
Страница 200: ......