2. Leakage Dose
Scope
IEC 60601-2-65 203.12.4
Requirements
In the LOADING STATE, the AIR KERMA due to LEAKAGE RADIATION from X-RAY
SOURCE ASSEMBLIES, 1 m from the FOCAL SPOT, average over any area of
100 cm
2
of which no principal linear dimension exceeds 20 cm, when operated at the
NOMINAL X-RAY TUBE VOLTAGE under condition of LOADING corresponding to the
reference LOADING conditions, shall not exceed 0.2 mGy in one hour.
Leakage Dose
Permissive Range
65 kVp, 3.0 mA, 0.5 s (Max. Exposure Condition)
At Focal Spot to Distance 1 m
Duty Cycle 1 : 60
< 0.2 mGy/h
Содержание EzRay Air W
Страница 1: ......
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...Notice vi EzRay Air W Model VEX S300W User Manual This page intentionally left blank...
Страница 8: ...Table of Contents viii EzRay Air W Model VEX S300W User Manual This page intentionally left blank...
Страница 18: ...2 Safety Instructions 10 EzRay Air W Model VEX S300W User Manual This page intentionally left blank...
Страница 26: ...3 System Overview 18 EzRay Air W Model VEX S300W User Manual This page intentionally left blank...
Страница 54: ...7 Disposing of the Unit 46 EzRay Air W Model VEX S300W User Manual This page intentionally left blank...
Страница 56: ...8 Product Specifications 48 EzRay Air W Model VEX S300W User Manual...
Страница 57: ...8 Product Specifications EzRay Air W Model VEX S300W User Manual 49 ENGLISH...
Страница 63: ...8 Product Specifications EzRay Air W Model VEX S300W User Manual 55 ENGLISH 5 Tube Dimensions mm...
Страница 66: ...8 Product Specifications 58 EzRay Air W Model VEX S300W User Manual This page intentionally left blank...
Страница 84: ......
Страница 85: ......
Страница 86: ......