10. Appendix
94
EzRay Air Portable
(Model: VEX-P300) User Manual
Immunity to
conducted
disturbances
induced IEC
61000-4-6
3 V 0.15-80 MHz
6 V in ISM bands
between 0.15 and
80 MHz
80% AM at 1 kHz
Power supply line
& I/O lines
3 V 0.15-80
MHz
6 V in ISM
bands between
0.15 and 80
MHz
80% AM at 1
kHz
Power supply
line & I/O lines
The strength of the RF
field in the frequency
range higher than 150
kHz ~ 80 MHz, the
strength of the RF field is
smaller than 3 V
The
magnetic
field of
supply
frequency
(50/60 Hz)
according to
IEC 61000-
4-8
30 A/m
30 A/m
The magnetic field of
power frequency should
coincide with the general
level found in the
business site or hospital
environment
Voltage dips
and
short
interruptions
IEC 61000-
4-11
0 %
U
T: 0.5 cycle
At 0°, 45°,
90°, 135°,
180°, 225°,
270° and
315°
0 %
U
T; 1 cycle
and
70 %
U
T; 25/30
Cycles Single
phase: at 0º
0 %
U
T; 250/300
cycle
0 %
U
T: 0.5
cycle
At 0°, 45°,
90°, 135°,
180°, 225°,
270° and
315°
0 %
U
T; 1 cycle
and
70 %
U
T; 25/30
Cycles Single
phase: at 0º
0 %
U
T;
250/300 cycle
The quality of supplied
power should be suitable
for the general business
site or hospital
environment.
If the user of the VEX-
P300 image intensifier
requires continued
operation during power
mains interruptions, it is
recommended that the
VEX- P300 image
intensifier is powered
from an uninterruptible
power supply.
Содержание EzRay Air VEX-P300
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...Notice vi EzRay Air Portable Model VEX P300 User Manual This page is intentionally left blank...
Страница 50: ...4 Operation 42 EzRay Air Portable Model VEX P300 User Manual This page is intentionally left blank...
Страница 66: ...5 Service Mode 58 EzRay Air Portable Model VEX P300 User Manual This page is intentionally left blank...
Страница 70: ...6 Troubleshooting 62 EzRay Air Portable Model VEX P300 User Manual This page is intentionally left blank...
Страница 76: ...8 Disposing of the Unit 68 EzRay Air Portable Model VEX P300 User Manual This page is intentionally left blank...
Страница 80: ...9 Device Specifications 72 EzRay Air Portable Model VEX P300 User Manual 4 Tube Dimensions mm...