Guidance and Manufacturer’s Declaration - Emissions  

The ClariVein

®

 IC

 

is intended for use in the electromagnetic environment specified below. The user ensure that it 

is used in such an environment.  

Emissions Test  

Compliance  

  

Electromagnetic Environment – Guidance  
  

RF Emissions CISPR 11  

Group 1  

The ClariVein

® 

IC uses RF energy only for its internal 

function. Therefore, its RF emissions are very low and 
are not likely to cause any interference in nearby 
electronic equipment.  

RF Emissions CISPR 11  

Class B  

 

Harmonics IEC 61000-3-2 

N/A 

 

Flicker IEC 61000-3-3 

N/A 

 

 

 

The ClariVein

®

 IC is suitable for use in all 

establishments, including domestic, and those directly 
connected to the public low-voltage power supply 
network that supplies buildings used for domestic 
purposes. 

 

Guidance and Manufacturer’s Declaration – Immunity  

The ClariVein

®

 IC is intended for use in the electromagnetic environment specified below. The user 

should ensure that it is used in such an environment.  

Immunity Test  

IEC 60601 Test 
Level  

Compliance 
Level  

Electromagnetic Environment – 
Guidance  

ESD

IEC 61000-4-2  

±6kV Contact 
±8kV Air  

±6kV 
Contact 
±8kV Air  

Floors should be wood, concrete or ceramic 
tile. If floors are synthetic, the r/h should 
be at least 30%  

EFT  

IEC 61000-4-4 

 

±2kV Mains  

±1kV I/Os 

 

N/A 

 

Mains power quality should be that of a 
typical commercial or hospital 
environment. 

 

Surge  
IEC 61000-4-5 

 

±1kV Differential  
±2kV Common 

 

N/A 

 

Mains power quality should be that of a 
typical commercial or hospital 
environment. 

 

Voltage Dips/Dropout  
IEC 61000-4-11 

 

>95% Dip for  
0.5 Cycle  
60% Dip for  
5 Cycles  
30% Dip for  
25 Cycles  
>95% Dip for  
5 Seconds 

 

N/A 

 

Mains power quality should be that of a 
typical commercial or hospital 
environment. If the user of the ClariVein

®

 

IC requires continued operation during 
power mains interruptions, it is 
recommended that the ClariVein

®

 IC be 

powered from an uninterruptible power 
supply or battery. 

 

Power Frequency  
50/60Hz  
Magnetic Field  

IEC 61000-4-8 

 

3A/m 

 

3A/m 

 

Power frequency magnetic fields should be 
that of a typical commercial or hospital 

 

 
 

____________________________________________________________________ 

ClariVein

®

 is a registered trademark of Vascular Insights

, LLC in the United States.    

           Page 9 

 

 

Содержание CLARIVEIN IC

Страница 1: ...is a specialty infusion catheter with 360 Rotatable Dispersion Wire connected to a proximally located integral battery powered Motor Drive Unit MDU The MDU includes the Speed Selector handle grip and...

Страница 2: ...ices and solutions intended for use including warnings cautions potential side effects and contraindications Before using ClariVein IC verify proper function and integrity of the device Refer to packa...

Страница 3: ...ral vasculature infusion procedure using the ClariVein IC are similar to those associated with any interventional procedure and include but are not limited to the following 1 Abrupt thrombosis and occ...

Страница 4: ...hnique to carefully remove the tray from the pouch and the contents from the tray 3 Inspect the device to be certain there are no visible signs of damage Do not use if contents are damaged 4 Remove Ba...

Страница 5: ...ve Unit 7 Mating Alignment Channel 11 Mating Alignment Channel Position 1 Position 2 8 Speed Selector 12 Battery Terminal Insulator Tab 9 Syringe Locking Support 13 Green Indicator Light 10 Trigger __...

Страница 6: ...Once placed into Position 2 the Catheter Assembly is not removable from the MDU 3 Using vascular imaging guidance confirm that the Dispersion Wire Tip remains in desired position Adjust as needed 4 Pl...

Страница 7: ...Wire is unsheathed Procedural Steps 1 Slowly withdraw the device through the treatment area while simultaneously infusing the physician specified agent Recommended technique o Depress the syringe plu...

Страница 8: ...tient Disable the MDU a Disengage the syringe from the Syringe Locking Support on the MDU Figure 2 9 b Rotate the catheter assembly to unlock from Position 2 Figure 4 c Move the Catheter Assembly Guid...

Страница 9: ...Level Compliance Level Electromagnetic Environment Guidance ESD IEC 61000 4 2 6kV Contact 8kV Air 6kV Contact 8kV Air Floors should be wood concrete or ceramic tile If floors are synthetic the r h sh...

Страница 10: ...r can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF Communications Equipment and the ClariVein IC as recommended below according to the max...

Страница 11: ...ility has been foreseen or was foreseeable even if Vascular Insights has been advised of the possibility thereof if and to the extent that Vascular Insights notwithstanding the foregoing is liable for...

Страница 12: ...Sterilized utilizing EO Consult instructions for use Keep away from rain BF Applied part Fragile handle with care Do not resterilize Do not use if package damaged Do not reuse ________________________...

Страница 13: ...ghts com US Patents 7 862 575 7 967 834 8 465 508 8 696 645 and other United States and foreign patents pending 2016 Vascular Insights LLC All rights reserved IFU 001 Rev B ___________________________...

Отзывы: