Guidance and Manufacturer’s Declaration – Immunity for ME Equipment/Systems that are NOT life 
supporting. 

The ClariVein

®

 IC is intended for use in the electromagnetic environment specified below. The user should 

ensure that it is used in such an environment.  

Immunity Test  

IEC 60601 
Test Level  

Compliance 
Level  

Electromagnetic Environment – Guidance  

Radiated RF  

IEC 61000-4-3  

3 V/m

80 MHz 

to 2.5 GHz  

(E1)=3V/m  

Portable and mobile communications equipment 
should be separated from the ClariVein

®

 IC by 

no less than the distances calculated/listed 
below:  
D=(3.5/V1)(Sqrt P) 150 to 80 MHz  

D=(3.5/E1)(Sqrt P) 80 to 800 MHz  
D=(7/E1)(Sqrt P) 800 MHz to 2.5 GHz  

where P is the max power in watts and D is the 
recommended separation distance in meters.  

Field strengths from fixed transmitters, as 
determined by an electromagnetic site survey, 
should be less than the compliance levels (E1).  
Interference may occur in the vicinity of 
equipment containing a transmitter.  

 
 

Recommended Separation Distances between portable and mobile RF Communications equipment and the 
ClariVein® IC which is not NOT Life-supporting. 

The

 

ClariVein

®

 IC is intended for use in the electromagnetic environment in which radiated disturbances are 

controlled. The user can help prevent electromagnetic interference by maintaining a minimum distance between 
portable and mobile RF Communications Equipment and the ClariVein

®

 IC as recommended below, according to 

the maximum output power of the communications equipment.  

Max Output Power 
(Watts)  

Separation (m)  
150 to 80MHz  

D=(3.5/V1)(Sqrt P)  

Separation (m)  
180 to 800MHz  

D=(3.5/E1)(Sqrt P) 

Separation (m)  
800MHz to 2.5GHz  
D=(7/E1)(Sqrt P

)

 

0.01  

0.116667  

0.116667  

0.233333  

0.1  

0.368932  

0.368932  

0.737865  

1  

1.166667  

1.166667  

2.333333  

10  

3.689324  

3.689324  

7.378648  

100  

11.66667  

11.66667  

23.33333  

 
WARRANTY  

Vascular Insights

, LLC (“Vascular Insights”) warrants that reasonable care has been used in the 

design and manufacture of this instrument and that it is free from manufacturing defects resulting 
from inferior workmanship or material that interferes with the ordinary use or functionality of the 
instrument for which it was sold by Vascular Insights

  or its authorized vendor. The warranty 

covers only manufacturing defects that are reported to Vascular Insights

 within one (1) year from 

the date that the instrument was delivered by Vascular Insights

 or its authorized vendor to the end 

user who purchased the instrument for his/her/its or his/her/its employees’ use. This warranty is the 

____________________________________________________________________ 

ClariVein

®

 is a registered trademark of Vascular Insights

, LLC in the United States.    

           Page 10 

 

 

Содержание CLARIVEIN IC

Страница 1: ...is a specialty infusion catheter with 360 Rotatable Dispersion Wire connected to a proximally located integral battery powered Motor Drive Unit MDU The MDU includes the Speed Selector handle grip and...

Страница 2: ...ices and solutions intended for use including warnings cautions potential side effects and contraindications Before using ClariVein IC verify proper function and integrity of the device Refer to packa...

Страница 3: ...ral vasculature infusion procedure using the ClariVein IC are similar to those associated with any interventional procedure and include but are not limited to the following 1 Abrupt thrombosis and occ...

Страница 4: ...hnique to carefully remove the tray from the pouch and the contents from the tray 3 Inspect the device to be certain there are no visible signs of damage Do not use if contents are damaged 4 Remove Ba...

Страница 5: ...ve Unit 7 Mating Alignment Channel 11 Mating Alignment Channel Position 1 Position 2 8 Speed Selector 12 Battery Terminal Insulator Tab 9 Syringe Locking Support 13 Green Indicator Light 10 Trigger __...

Страница 6: ...Once placed into Position 2 the Catheter Assembly is not removable from the MDU 3 Using vascular imaging guidance confirm that the Dispersion Wire Tip remains in desired position Adjust as needed 4 Pl...

Страница 7: ...Wire is unsheathed Procedural Steps 1 Slowly withdraw the device through the treatment area while simultaneously infusing the physician specified agent Recommended technique o Depress the syringe plu...

Страница 8: ...tient Disable the MDU a Disengage the syringe from the Syringe Locking Support on the MDU Figure 2 9 b Rotate the catheter assembly to unlock from Position 2 Figure 4 c Move the Catheter Assembly Guid...

Страница 9: ...Level Compliance Level Electromagnetic Environment Guidance ESD IEC 61000 4 2 6kV Contact 8kV Air 6kV Contact 8kV Air Floors should be wood concrete or ceramic tile If floors are synthetic the r h sh...

Страница 10: ...r can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF Communications Equipment and the ClariVein IC as recommended below according to the max...

Страница 11: ...ility has been foreseen or was foreseeable even if Vascular Insights has been advised of the possibility thereof if and to the extent that Vascular Insights notwithstanding the foregoing is liable for...

Страница 12: ...Sterilized utilizing EO Consult instructions for use Keep away from rain BF Applied part Fragile handle with care Do not resterilize Do not use if package damaged Do not reuse ________________________...

Страница 13: ...ghts com US Patents 7 862 575 7 967 834 8 465 508 8 696 645 and other United States and foreign patents pending 2016 Vascular Insights LLC All rights reserved IFU 001 Rev B ___________________________...

Отзывы: