TECHNICAL INFORMATION
N. Read/
Write
Data
Type
Description Admitted
Values
N. Read/
Write
Data
Type
Description Admitted
Values
121 R/W N Maximum
rota-
tional frequency
in Hz (active only
in Stop condition)
150 to 963
(default = 963)
122
R/W
L
Set vent valve
on/off
(on = closed)
On = 1
Off = 0
(default = 1)
123
124
Reserved to Varian service
125
R/W
L
Set the vent valve
operation
Automatic = 0
(see note 1.)
On command = 1
(see note 2.)
126 R/W N Vent
valve
open-
ing
delay
(expressed in 0.2
sec)
0 to 65535
(corresponding to
0 to 13107 sec)
130
Reserved to Varian service
200 R
N Pump
current in mA dc
201
R
N
Pump voltage in
Vdc
202
R
N
Pump power in
W (pump current
x pump voltage
duty cycle)
203 R N Driving
frequency
in Hz
204 R N Pump
tempera-
ture in °C
0 to 70
205
R
N
Pump status
Stop = 0
Waiting intlk = 1
Starting = 2
Auto-tuning = 3
Braking = 4
Normal = 5
Fail = 6
206
R
N
Error code
Bit description:
see the following
figure
210 R
(see
note 3.)
N Actual
rotation
speed in Hz
100 to 963
300
R
N
Cycle time in
minutes (zeroed
by the reset com-
mand)
0 to 999999
301 R N Cycle
number
(zeroed by the
reset command)
0 to 9999
302
R
N
Pump life in
hours (zeroed by
the reset com-
mand)
0 to 999999
320
to
399
Reserved to Varian service
400 R A CRC
EPROM
(QE)
QE8XXXX
(where “XXXX”
are variable)
402 R A CRC
Param.
(PA)
PA8XXXX
(where “XXXX”
are variable)
404 R A CRC
Parameter
structure
"XXXX"
500
Reserved to Varian service
503
R/W
N
RS 485 address
0 to 31
(default = 0)
504
R/W
L
Serial type se-
lect
0 = RS 232
1 = RS 485
(default = 0)
NOTES
1. Automatic means that when the controller
stops, the vent valve is opened with a delay de-
fined by window n. 126; when the controller
starts, the vent valve is immediately closed.
2. On command means that the vent valve is
opened or closed by means of window n. 122.
3. These commands remove the pump motor
supply for a short time. If they are used con-
tinuously, they can cause the pump stopping.
7 6 5 4 3 2 1 0
TOO
HIGH
LOAD
NO
CONNECTION
SHORT
CIRCUIT
PUMP
OVERTEMP.
OVERVOLTAGE
CONTROLL.
OVERTEMP.
AUX FAIL
POWER FAIL
Window N. 206 Bit Description
80