ProStar 701 Fraction Collector Control Manual
- 27 -
03-914947-27:2
: moves to the following tube.
: moves to the next rack and keeps the same tube number position.
: allows the user to clear the synopsis screen (displayed in the
Overview tab). User has to activate this button between two
acquisitions to reset the rack (if he replaces tubes containing
fractions by empty ones); if not, the fractions of the second
acquisition will be put at the end of the first acquisition ones.
: activating this button stops the fraction collection. User can stop
and start again the collection during an acquisition, the log will be
created according these stop/start actions. Note that when
performing collection under Galaxie control, the program A of the
ProStar 701 Fraction Collector is erased and replaced by the one
defined in the Galaxie method.
Overview Window
Click on the
button: the
Status
Overview
window is
displayed and contains three tabs:
General:
displays the synopsis of the instrument, where
some current values can be viewed: rack filling in, racks
properties.
If moving the cursor into a rack, the tube position is
mentioned in the left bottom corner of the first rack.