ProStar 500 Column Valve Module Manual
- 16 -
03-914947-57:1
require the heater block to come within 2
C of that temperature
and stabilize for that delay time.
In the
Valve X delay
field, setting the valve delay will delay the
valve from activating at its programmed time by the delay value.
With a delay of 1 minute, a valve switch programmed at 5
minutes will occur at 6 minutes.
Miscellaneous parameter group:
In the
Transition delay
field, enter the time to wait before
injecting after all prerun settings should be applied.
Enter the
End Time
of the valve program which table is below.
After the end time, all programmed changes will not be applied.
Enter the post run time to be applied in the
Equilibration Time
field.
Choose
Reset
in the
Action field
to apply to each valve and
heater the prerun settings at the end of the run. Choose
Hold
to
keep the current valves and heaters parameters at the end of
the run.
Use the table to build a column, solvent and temperature
program.
Use the following buttons to edit the table:
to add a line to the table.
to delete a line in the table.
to clear the entire table.
In the table, enter the time when the change must occur in the
Time
field. In the
ValveX
and
HeaterX
fields, choose the
column names, the solvent names and the temperatures that
must be applied at the entered time. Only the column and
solvent defined in the configuration are available.