TECHNICAL INFORMATION
5
87-900-847-01 (B)
HAND-HELD TERMINAL DESCRIPTION
The Hand-held terminal is on LCD alphanumeric
display and keyboard, which provides useful real time
indications of the operating conditions/parameters of
the Turbo-V 60 pump series when connected to a
Turbo-V 60 controller without front panel.
Moreover, it is possible to reprogram the controller
parameters or troubleshooting pump/controller.
The Hand-held terminal, 188 mm height x 110 mm wide
x 41 mm deep (7.4" x 4.3" x 1.6"), is connected and
powered to the Turbo-V controller via a 1.5 meters (5 ft)
cable.
USE
General
Make all vacuum manifold and electrical connections and
refer to Turbo-V pump instruction manual before to
operating the Turbo-V controller.
!
To avoid injury to personnel and damage to the
equipment, if the pump is laying on a table make sure it
is steady. Never operate the Turbo-V pump if the pump
inlet is not connected to the system or blanked off.
Connect the hand-held terminal to the Turbo-V controller.
Plug the controller power cable into a suitable power
source.
If the hand-held terminal is used to monitor the pump
operating conditions, follow paragraph "Monitor Mode";
if reprogramming or trouble shooting is required, follow
paragraph "Program Mode"; and the following
paragraph.
Monitor Mode
By pressing the CURRENT push-button, the display
shows:
I
=
0
.
0
0 A
P
=
0 W
X
.
.
K R P M
X X
°
C
where:
•
I
= is the DC current drawn by the pump range
(0.00 to 9.99 Ampere)
•
P
= is the DC power drawn by the pump (range 0 to
999 Watt)
•
KRPM
= is the theoretical rotational speed of the
pump as a function of the controller output
frequency (range 17 to 99 KRPM)
•
°C
= is the temperature of the outer ring of the
upper bearing (range 00 to 99 °C)
•
X
= during operation a selected set point condition
(1 or 2 contrast inverted) appears when the
programmed threshold speed value is not reached.
•
Press the CYCLE NUMBER once and the display
shows:
X X X X C Y C L
E
X X X X X m
P U M P
L
I
F
E
X X X X X h
where:
•
CYCLE
= are the cycles performed (range 0 to 9999)
•
m
= is the elapsed time related to the cycle number
displayed (range 0 to 99999 minutes)
•
PUMP LIFE
= is the total operation time of the pump
(range 0 to 99999 hours).
Program Mode
Refer to paragraph "Speed Adjustment" to program/
adjust the high and low speed values. This feature is
only applicable when a dedicated software is installed
on the Turbo-V 60 controller microprocessor (e.g. for
leak detector operation).
FRONT / REMOTE/ 232 Selection
•
Press CYCLE NUMBER and PUMP CURRENT
push-buttons together for at least 2 seconds and
the processor enters in a routine where it is
possible to program the controller.
In this routine, the CYCLE push-button is used for
choosing/changing the value or condition; the
PUMP CURRENT push-button is used to enter and
confirm the value. At any time it is possible to exit
this routine by pressing the CYCLE and PUMP
CURRENT push-buttons at the same time for at
least 2 seconds.
The display shows:
F R O N T
/
R E M O T E
/
2
3
2
S E L E C T
I
O N
:
X X X X X X
where:
XXXXXX
= means the word FRONT or
REMOTE, or RS 232 depending on the last
selection.
Select the FRONT panel operation if the front panel
command is used.
After choosing the desired selection by pressing the
CYCLE push-button, press the PUMP CURRENT
push-button to enter the value.
WARNING!