SECTION II - INSTALLATION

87-900-879-01 (A)

4

2-1 General

Inspect the controller for any shipping damage.

The controller is designed to be installed into a rack.

All the connections are fitted to J2, J3, J4 and J5
connectors.

Should the controller be connected to a host computer
via the-RS-232 interface, a suitable cable must be
prepared.

In the following paragraphs are detailed the
input/output signals.

NOTE

The PCB installed into the customer system must be
positioned so that cold air (forced or natural
convection) can flow through the PCB components.

2-2 J2 connector

The signals of J2 connector are the following:

  Pins 1/2 Power supply 24.  Vdc (20 to 32 Vdc) (pin

2 positive).

  Pins 4/3 Kl relay contacts. The relay switches when

the output frequency overrides the set threshold
(800 Hz). The threshold overriding is indicated also
on pin 6 of J4 connector.

2-3 J3 connector

The signals of J3 connector are the following:

  Pin 1   42 Vac 3-phase output to pump motor stator

(phase T).

  Pin 2   42 Vac 3-phase output to. pump motor stator

(phase S).

  Pin 3   42 Vac 3-phase output to pump motor stator

(phase R).

  Pins 5/4   Pump temperature sensor.

2-4 J4 connector

The signals of J4 connector are the following:

  Pin 3 Failure. When a fault condition is detected,

the pump is stopped and a signal is present on this
pin.

  Pin 5 Low Speed mode activated. When the Low

Speed mode is activated a signal is present on this
pin.

  Pin 6 When the output frequency is less than 800

Hz, a signal is present on this pin.

  Pin 7 Analog output. A voltage from 0 to 10 Vdc

proportional to the supply current (from 0 to 2.5 A).

 

  Pin 8 Analog output. A voltage from 0 to 10 Vdc

proportional to the output pump motor driving
frequency (from 0 to 1250 Hz).

 

  Pin 11 Error reset command. Should an operation

fault be detected (presence of signal on pin 3 of J4),
the pump is stopped. To reset the fault you must
connect pin I I to ground (pin 15 of J4) for at least I
sec.

 

  Pin 12 ON/OFF pump.  To stop the pump you must

connect pin 12 to ground (pin 15 of J4); to start the
pump you must remove the contact between pin 12
and ground.

 

  Pin 13 Low Speed mode activation.  When the

contact between pin 13 and ground (pin 15 of J4)
closes, the turbopump runs at low speed and when
the contact opens, the turbopump reverts to high
speed mode. When the Low Speed mode is
activated an output signal is present on pin 5 of J4.
The Low Speed frequency is 830 Hz.

2-5 J5 RS-232 interface connector

The signal of J5, the RS-232 interface connector, are
the following:

  Pin 1 Ground.

  Pin 2 RXD. Received data.

  Pin 3 TXD. Transmitted data.

Содержание 969-9514

Страница 1: ...87 900 879 01 A APRIL 2002 INSTRUCTION MANUAL Turbo V70 controller Eurocard Model 969 9514 vacuum technologies...

Страница 2: ...utline 3 II Installation 2 1 General 4 2 2 J2 connector 4 2 3 J3 connector 4 2 4 J4 connector 4 2 5 J5 RS 232 interface connector 4 2 6 RS 232 communication description 5 III Operation 3 1 General 7 3...

Страница 3: ......

Страница 4: ......

Страница 5: ...operation maintenance manuals and any additional information provided by Varian All warnings and cautions should be read carefully and strictly observed Address any safety operation and or maintenance...

Страница 6: ...ncy converter which is driven by a single chip microcomputer and is composed of a PCB which includes all the circuitry necessary for its operation The microcomputer generates the variable output volta...

Страница 7: ...SECTION I DESCRIPTION 87 900 879 01 A 3 1 4 Controller outline The outline dimension for the controller are shown in Fig 1 1 Figure 1 1 Controller outline...

Страница 8: ...the following Pin 3 Failure When a fault condition is detected the pump is stopped and a signal is present on this pin Pin 5 Low Speed mode activated When the Low Speed mode is activated a signal is p...

Страница 9: ...ter reading request 06 parameter writing Data in this controller it can be in the case of parameter reading request 000000 or 111111 in the case of YES NO command or parameter writing a six numeric ch...

Страница 10: ...l to 1 can be only read equal to 1 can be only written PARAMETER NUMBER TYPE DESCRIPTION X0002 D Low speed ON OFF X0003 D Pump ON OFF X0008 D Remote ON OFF 10009 D Error acknowledgment error status cl...

Страница 11: ...the same connector To switch off the pump is necessary to connect pin 12 of J4 connector to pin IS ground of the same connector 3 3 Low speed activation deactivation To activate the Low Speed status...

Страница 12: ...signals the error occurred by means of a diagnostic LED located on the board and on the RS 232 port The LED blinks in a coded mode it flashes a number of time equal to the error code see the following...

Страница 13: ...TIFICATION Product Description Varian P N Varian S N Purchase Reference TYPE OF RETURN check appropriate box Paid Exchange Paid Repair Warranty Exchange Warranty Repair Loaner Return Credit Shipping E...

Страница 14: ...rate No zero high backround Gauge tube not working Display problem Vacuum system unstable Cannot reach test mode Communication failure Degas not working Failed to start Other Error code on display Oth...

Страница 15: ...an India PVT LTD 101 108 1st Floor 1010 Competent House 7 Nangal Raya Business Centre New Delhi 110 046 India Tel 91 11 5548444 Fax 91 11 5548445 Italy Varian Vacuum Technologies Via F lli Varian 54 1...

Отзывы: