7.
Technical specification
Classifications
See section 2.11
Application
Mobile and stationary equipment for radiography (see §2.1)
Materials
ABS, PC, POM, stainless steel, aluminium, lead
Power source
VDC Models
24 VDC ± 10%, 24 VA
Optional, depending on model
VDC / VAC Models
12-45 VDC ± 10% / 24 VA
20-30 VAC ± 10% / 35 VA 50/60 Hz
Application
Limiting primary X-ray beam, part of an X-ray system
Light field luminosity
≥ 160 lx
Light field / X-ray alignment
≤ 1% SID
Light field edge contrast ratio
≥4
SID tape measure range
90 – 200 [cm]
Field size
Min. 3x3cm @ 1m SID [1.2”x1.2” @ 39.4” SID]
Max. 43x43cm @ 1m SID [16.9”x16.9” @ 39.4” SID]
Weight
±8 kg
Size
(WxDxH) 230x267x177mm [9.1”x10.5”x7.0”]
Leakage
≤ 0.5 mGy/h
Filter module (added filtration)
(optional feature)
1 mm Al + 0.1 mm Cu
1 mm Al + 0.2 mm Cu
2 mm Al
None
Operation environment
Ambient temperature: 15 – 40 °C
Relative humidity:
30 – 75 %, non-condensing
Atm. Pressure:
750 – 1030 hPa
Storage
Ambient temperature: -20 – 60 °C
Relative humidity:
30 – 95 %, non-condensing
Atm. Pressure:
750 – 1030 hPa
Maximum cable length
< 30 m (applies to all cables)
Cable type(s)
Unshielded or shielded; Can-bus H/L signals preferably twisted