5. Power and Environmental Specifications
Table 5: Computer environmental conditions
Temperature
Relative humidity
Component
Transport
and storage
Operation
Transport
and storage
Operation
Computer
-20–55°C
10-32°C
10–90%
20–75%
Table 6: Detector environmental conditions
Temperature
Relative humidity
Air Pressure
Component
Transport
and storage
Operation
Transport
and storage
Operation
Operation
4343Rv3/RC
-20-70°C
10-35°C
10-90%
10-90%
700-1060hPa
4336Wv4
-20-70°C
10-35°C
10-90%
10-90%
700-1060hPa
4343W
-20-55°C
10-40°C
10-90%
10-90%
700-1060hPa
4343W with
optional tether
-20-55°C
10-40°C
10-90%
10-90%
700-1060hPa
4336W-G5
-20-55°C
10-40°C
10-90%
10-90%
700-1060hPa
Electromagnetic compatibility
The system complies with IEC 60601-1-2 ed4.1. Prevent the potential risk of electromagnetic
interference between this equipment and other devices.
The system is intended for use in the electromagnetic environment in which radiated
disturbances are controlled. The customer or user of the system can help to prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment and the system as recommended below, according to the
maximum output power of the communications equipment.
The system has been tested for electromagnetic compatibility (EMC) compliance, but
interference can still occur in an electromagnetically noisy environment. Maintain a suitable
distance between electrical devices to prevent cross-interference. The PC cabinet should be
placed as far as possible from any device that generates large amounts of electromagnetic
disturbance.
CAUTION:
Medical electrical equipment requires special precautions
to maintain electromagnetic compatibility. The system must be
installed and put into service according to the EMC information
provided in this document. Portable and mobile RF communications
equipment can affect medical electrical equipment.
Warning:
Portable RF communications equipment (including
peripherals such as antenna cables and external antennas)
should be used no closer than 30 cm (12 inches) to any part of the
MODEL OR TYPE REFERENCE of the ME EQUIPMENT or ME
SYSTEM, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
Nexus DR Regulatory and Safety Manual, 736-226-G1, Rev. J | 25
Содержание nexus DR
Страница 1: ...Regulatory and Safety Manual 736 226 G1 Rev J...
Страница 2: ...Regulatory and Safety Manual 2 Nexus DR Regulatory and Safety Manual 736 226 G1 Rev J...
Страница 13: ...1 Regulatory Information 4 Nexus DR Regulatory and Safety Manual 736 226 G1 Rev J...
Страница 17: ...2 Statement of Intended Use 8 Nexus DR Regulatory and Safety Manual 736 226 G1 Rev J...
Страница 21: ...3 Cleaning Practices 12 Nexus DR Regulatory and Safety Manual 736 226 G1 Rev J...
Страница 41: ...5 Power and Environmental Specifications 32 Nexus DR Regulatory and Safety Manual 736 226 G1 Rev J...
Страница 44: ...Index 35...
Страница 45: ......