INDICATIONS, WARNINGS AND CAUTIONS

Primary Indications:

The Oxygen Assist Module is an optional module used only with the Vapotherm Precision Flow and 
is indicated for on-demand titration of oxygen into warm humidified breathing gases delivered to 
spontaneously breathing patients based on continuous non-invasive monitoring of blood oxygen 
saturation. 

The Oxygen Assist Module is intended to treat pediatric (including neonatal) and adult patients in 
monitored clinical environments.

Incorporation of SpO2 into the Precision Flow with the Oxygen Assist Module does not eliminate 
the need for separate and independent patient monitoring indicated by the Precision Flow system 
labeling. 
Patients on high velocity nasal insufflation receiving supplemental oxygen are acute and appropriate clinical 
vigilance should be observed by the care team. Additional patient monitoring including pulse oximetry is 
necessary if the Precision Flow is used to give supplementary oxygen.

Oxygen Assist Module Packaging contains:

Oxygen Assist Module Unit

Quick Reference Guide

Power Cord

Pole Mount Assembly

SpO2 Patient Cable

ALARM/ALERT MESSAGE

ACTION                                   

 %O

2

 Above Limit*

Increase %O

2

 Alarm Limit, or enter MANUAL mode

Battery Backup In Use

Oxygen Assist Module is in Battery Mode. When external power is lost, 

the module will automatically switch to Battery Mode

Battery Low

Plug the Oxygen Assist Module into external power

Check Precision Flow

Clear Precision Flow alarm and ensure it is in Run Mode with both Air 

and O

2

 sources present

Check SpO

2

*

See the alert message on screen for additional information. Reconnect 

sensor with good signal or replace SpO

2

 sensor or patient cable

Local Storage is Low

Delete Previous Case Files from Local Storage

Local Storage is Full

Delete Previous Case Files from Local Storage

SpO

2

 Target > 95%

Change SpO

2

 Target to 95% or Below

SpO

2

 Target < 85%

Change SpO

2

 Target to 85% or Above

Time Near Expiration

Insert new Time Card before time expires or End Case

SYSTEM SET-UP

•  Connect Oxygen Assist Module to Precision Flow and ensure the Precision Flow is in 

RUN mode and shows a GREEN LED light

•  Connect the SpO2 patient cable to the Oxygen Assist Module. Attach the sensor to the 

patient cable and apply the sensor to the patient

•  Plug Oxygen Assist Module into external power and power on

•  On the System screen, check settings

•  On the Patient screen, check settings. Select parameters with the touchscreen and 

rotate the blue Setting Control Knob to adjust the value.

•  Start Case

•  Press the O2 Mode AUTO button to begin automatic mode once the patient is in a 

stable SpO2 state.