 P/N VT-108723-US, Manual revision: C

                         

Vante

® 

SEBRA

® 

MINI

™ 

Tube Sealer Power Source

 

 

 

Table 5-1, Troubleshooting guide

 

 

Amber indicator light 
on battery charger 
fails to illuminate. 

Battery pack improperly 
connected to charger. 

Be sure that the battery 
pack is properly 
connected to the 
charger. 

Battery is dead. 

Replace battery. 

Battery charger 
improperly connected to 
AC power source. 

Be sure the battery 
charger is connected to 
the proper AC power 
source. 

 

Note: Arcing at the sealing head may cause a battery low condition to be 
indicated, even when the battery is not low.

 

 

Failure isolation

 

If the Sealer fails to operate properly, connect a different sealing head and RF 
power cable to the power source to isolate the failed component. 

If the Sealer performs properly, then the questionable sealing head is probably 
defective. If the Sealer fails to function properly, then the power source may be 
defective. In any event, contact the local Vante representative. 

 

 

C

USTOMER SERVICE

 

Services 

Clinical training

 

The local Vante representative will provide staff training upon delivery  of the 
equipment and should be contacted to organize further instruction if  needed. 

 

Repair service

 

Vante maintains a worldwide network of company-trained service 
representatives responsible for responding to technical needs concerning 
equipment. If service beyond the routine maintenance and cleaning described 
in this manual is required, the local Vante representative should be  contacted 
to provide specific instruction. 

 

Product return guidelines

 

If, for any reason, merchandise must be returned to the company, the customer 
should contact the local Vante representative to arrange for repairs or  returns 
using  procedures  to  ensure  proper  handling  and  subsequent  analysis.  No 
returns will be accepted without advanced authorization. 

 

Warning: Vante products must be properly cleaned and packaged  prior 
to their return. It remains an important responsibility of the  customer to 
reduce potential health hazards by being aware of the risks  involved in 
the shipping, handling and testing of this material. Units  returned to 
Vante for repair are subject to biohazard charges if  any component is 
contaminated with blood or blood products.

 

Содержание SEBRA MINI 2380

Страница 1: ...operation manual VT 108723 US MINI Model 2380 Hand Held Tube Sealer Power Source VANTE INC IGNITING INNOVATION...

Страница 2: ...ii 2007 2011 2017 Vante Inc All rights reserved...

Страница 3: ...aimer This manual is intended as a guide to provide the operator with necessary instructions on the proper use and maintenance of certain Vante Corporation products This manual should be used in conju...

Страница 4: ...Vante SEBRA MINI Tube Sealer Power Source P N VT 108723 US Manual revision C iv...

Страница 5: ...2 Product description Product overview 5 Application 5 Theory of operation 5 Component identification 7 Operating environment 9 PVC tubing size categories 10 Sealing modes 10 Duty cycle 11 Seal spaci...

Страница 6: ...mospheres 27 List of Figures Figure 2 1 2380 MINI Hand Held Tube Sealer Components 7 Figure 2 2 Model 2380 Power Interface Panel 7 Figure 2 3 Serial Plate for Model 2380 8 Figure 3 1 Opening Battery C...

Страница 7: ...handling of blood and or blood products The terms Note Caution and Warning are used in this manual with the following symbols to notify the operator of important and or critical information Note provi...

Страница 8: ...ls may be found on the device or device packaging Shock hazard Attention Consult accompanying documents Protective earth ground Identifies any terminal intended for connection to an external conductor...

Страница 9: ...m 1 6 in 12 2 cm 4 8 in Weight of sealing head 0 3 kg 0 6 lbs Weight of power source carryingpouch strap and battery pack 2 1 kg 4 7 lbs Table 1 2 Environmental Conditions Condition Value Ambient oper...

Страница 10: ...specifications Table 1 3 Electrical Specifications Parameter Specification Power source 12 V NiCad battery pack Battery charger Input Output 100 240VAC 50 60 Hz 15VDC 55mA and 1 4A RF power output 40...

Страница 11: ...blood products is typically made from thermoplastic vinyl namely polyvinylchloride or PVC and it is this application for which this product is specifically intended Theory of operation The physical p...

Страница 12: ...accommodated Additionally a built in arc detect mode will normally preclude sealing if the tubing exterior ground jaw and or RF jaw surfaces are affected by contamination and or moisture Thus the Mode...

Страница 13: ...I Hand Held Tube Sealer Components Shown in 2 2 is the power interface panel on the front of the RF power source The serial plate 2 3 located on the underside of the RF power source indicates the mode...

Страница 14: ...the RF jaw tube assembly ground jaw and lever Seals tubing by compressing it between the RF and ground jaws while applying RF energy Jaws must remain closed for one second after indicator light turns...

Страница 15: ...n the sealing head and tubing are clean and dry and to frequently check the quality of seal being produced The Power Source and Sealing Head should be stored between 10 and 122 F 12 and 50 C To ensure...

Страница 16: ...g modes Intermittent mode refers to a sealing process in which no more than 5 seals are produced rapidly as quickly as the Sealer can normally produce for the tubing size being used within a minimum 1...

Страница 17: ...in Chapter 3 of this manual Table 2 3 lists suggested duty cycles to help users estimate their maximum allowed sealing rate s Table 2 3 Suggested Duty Cycles Sealing Head Sealing Mode PVC Tubing Size...

Страница 18: ...ag tubing follow the instructions provided by the bag manufacturer General guidelines Seals that progress along an open ended length of tubing may be spaced as desired but inch 1 cm is the recommended...

Страница 19: ...rs and packing materials for future use 2 For optimal performance batteries should be fully charged prior to use 3 Check to see that the power switch is in the off position Open the battery compartmen...

Страница 20: ...underneath and ensuring the wires are aligned properly See 3 3 below Note Ensure the latch is underneath the device and position hand underneath the battery while inserting in the case This prevents...

Страница 21: ...and latch the battery compartment door If desired install the RF power source into its carrying pouch Caution Do not use the battery outside the sealer case or operate the sealer without a door Damag...

Страница 22: ...he RF ready light on the power panel will illuminate which indicates that the Sealer is ready for use 10 See the Operation Manual for the Sealing Head for information on handling and sealing with the...

Страница 23: ...e power source off and be sure the sealing head and or tubing are dry and clean to eliminate the possibility of further arcing Turn the power source on and continue sealing If the battery low conditio...

Страница 24: ...ery should slide out of device when removing Figure 4 1 Properly Removing the Battery Using the Strap Device is held at a downward angle Battery should slide out of device when removing Figure 4 2 Pro...

Страница 25: ...charger connectors are keyed to properly engage Be sure that the RED battery pack connector is engaged into the RED battery charger connector and the BLACK into the BLACK If they do not easily engage...

Страница 26: ...arge period amber charge status light is continuously illuminated exceeds 6 hours disconnect battery pack from battery charger and contact Vante s Customer Care Center Caution If Pending Mode period a...

Страница 27: ...d between the ceramic RF jaw and the ground jaw General cleaning of the sealing head see the Operation Manual for the Sealing Head should be performed periodically or when blood or moisture contaminat...

Страница 28: ...ration Manual for the Sealing Head Battery is dead Replace battery Power Source Malfunction Call Vante Hand Held Sealing Head malfunction See Operation Manual for Hand Held Sealing Head RF cable malfu...

Страница 29: ...local Vante representative will provide staff training upon delivery of the equipment and should be contacted to organize further instruction if needed Repair service Vante maintains a worldwide netw...

Страница 30: ...P N VT 108723 US Manual revision C Vante SEBRA MINI Tube Sealer Power Source 24...

Страница 31: ...s RF effects on pacemakers There is no evidence that Vante RF instruments interfere with the function of modern cardiac pacemakers Electrical safety Vante RF devices meet or exceed appropriate electri...

Страница 32: ...Vante Inc 3480 E Britannia Dr Suite 120 Tucson AZ 85706 USA www vante com...

Страница 33: ...27 P N VT 108723 US Manual revision C Vante SEBRA MINI Tube Sealer Power Source 2007 2011 2017 Vante Inc...

Отзывы: